Impella Purge Solution Requirements
A single D5W solution with 25 mEq sodium bicarbonate can be used for each individual Impella device—you need one purge solution per pump, not a shared solution for multiple devices.
Understanding Impella Purge Solution Setup
Each Impella device requires its own dedicated purge solution system. The purge fluid runs continuously through the device to prevent biomaterial deposition and maintain pump function. You cannot share a single purge solution bag between multiple Impella devices.
Standard Purge Solution Composition
The bicarbonate-based purge solution (BBPS) consists of 1:
- D5W (5% dextrose in water) as the base solution
- Sodium bicarbonate 25 mEq/L as the additive (can range from 25-50 mEq/L)
- This creates a hypertonic solution with pH approximately 8.0
When to Use Bicarbonate-Based Purge Solution
BBPS is FDA-approved as an alternative to heparin-based purge solutions for patients who 2:
- Have heparin intolerance or contraindications
- Develop heparin-induced thrombocytopenia (HIT)
- Experience major bleeding complications
- Have significant coagulopathy
The bicarbonate solution works by increasing protein stability, reducing protein denaturation at the radial gap, and decreasing blood coagulation at the axial gap 3.
Clinical Evidence and Duration of Use
Multiple case series demonstrate successful BBPS use:
- Duration: Patients have been successfully supported for 11-21+ days on BBPS 2, 4
- Device function: No significant changes in purge pressure, flow, or motor current when properly managed 4
- Complications: In one series of 18 patients, only 17% experienced Impella complications, with two occurring after 21 days of therapy 2
Important Caveats
Monitor closely for device complications, particularly with prolonged use beyond 21 days. Three key parameters to track:
- Purge pressure (should remain stable)
- Purge flow rate (should remain consistent)
- Motor current spikes (indicating potential biomaterial deposition)
Each Impella device requires approximately 30-60 mL/hour of purge solution, so calculate your D5W + sodium bicarbonate preparation accordingly for 24-hour periods 5, 6.
Special Considerations
For patients requiring PET imaging, a sterile water-based bicarbonate purge solution (without dextrose) can be used temporarily during the scan to avoid interference with imaging 6. This requires close monitoring but has been successfully performed without adverse pump effects.
The bicarbonate concentration can be adjusted between 25-50 mEq/L based on clinical needs, with higher concentrations showing enhanced protein stability and anticoagulant effects 3. However, avoid extremes of alkalemia (pH should not exceed 7.50-7.55) to prevent iatrogenic harm 7.