How to Use Jackson-Rees for Neonatal Resuscitation to Provide PEEP
When using a Jackson-Rees (Mapleson-E/anesthesia-type) circuit for neonatal resuscitation, you must attach a PEEP valve to the tail-piece and manually occlude the tail opening to approximately 5 cm H₂O to deliver positive end-expiratory pressure, though this requires significant skill and is less reliable than purpose-built devices. 1
Device Characteristics and PEEP Delivery
The Jackson-Rees circuit is a flow-inflating bag system that:
- Requires a compressed gas source to function (unlike self-inflating bags) 1
- Can deliver PEEP when properly configured with a flow-control tail-piece and Norman elbow 2
- Demands more operator skill than other devices to use effectively 1
Technical Setup for PEEP
To provide PEEP with a Jackson-Rees circuit:
- Attach a flow-control tail-piece to the reservoir bag 2
- Partially occlude the tail opening during ventilation to maintain back-pressure
- Target approximately 5 cm H₂O PEEP for preterm infants 1
- Monitor pressure continuously if possible, as manual control is inconsistent 2
Critical Limitations and Pitfalls
The major problem with Jackson-Rees circuits is inconsistent PEEP delivery. Research demonstrates that:
- Only experienced respiratory therapists could reliably maintain target PEEP pressures with anesthesia-type bags 2
- Non-therapist operators (nurses, physicians, residents) produced highly variable PEEP levels (median 2.9 ± 1.2 cm H₂O with Jackson-Rees versus target 5 cm H₂O) 2
- Pressure fluctuates significantly during prolonged inflations—the difference between first and fifth second averaged 7.1 cm H₂O with anesthesia bags 2
Evidence-Based Recommendations
When PEEP Should Be Used
For preterm newborns requiring positive-pressure ventilation, use approximately 5 cm H₂O PEEP (Class IIb recommendation) 1. This recommendation applies specifically to:
- Preterm infants <35 weeks gestation who are apneic or bradycardic 1
- Any infant requiring PPV to establish functional residual capacity 1
The evidence shows PEEP may slightly reduce supplemental oxygen requirements, though it doesn't significantly impact mortality, bronchopulmonary dysplasia, or other major outcomes 1. For term infants, insufficient data exists to make a recommendation 3.
Device Selection Hierarchy
Given the limitations of Jackson-Rees circuits, consider alternative devices when available:
- T-piece resuscitator (Neopuff): Most consistent PEEP delivery across all operator skill levels—maintains target pressures within 0.2 cm H₂O variation 2
- Flow-inflating bag with PEEP valve: Reliable PEEP but requires practice 1
- Self-inflating bag with PEEP valve: Cannot deliver CPAP and may not achieve PEEP reliably even with valve attached 1
- Jackson-Rees circuit: Functional but operator-dependent 2
Practical Algorithm for Jackson-Rees Use
If you must use a Jackson-Rees circuit:
- Set gas flow rate at 10-15 L/min to maintain bag inflation
- Achieve proper mask seal before attempting ventilation
- Squeeze bag to deliver breath at target peak pressure (20-40 cm H₂O for initial breaths) 4
- Partially occlude tail-piece during exhalation phase to maintain 5 cm H₂O back-pressure
- Watch for chest rise as primary indicator of effective ventilation
- Monitor heart rate response—should increase to >100/min within 30 seconds 1
Common pitfall: Operators often fail to maintain consistent tail-piece occlusion, resulting in zero PEEP despite having the equipment 2. This is why respiratory therapists achieve better results—they have more experience with manual pressure control 2.
Context and Nuances
The 2015 AHA guidelines acknowledge that PEEP evidence is weak (Class IIb, Level of Evidence B-R), with randomized trials showing no improvement in critical outcomes 1. However, the recommendation persists based on:
- Physiologic rationale from NICU ventilation experience 1
- Potential benefit in establishing functional residual capacity 1
- Low risk of harm when properly applied 1
The 2024 meta-analysis suggests PEEP reduces mortality in preterm infants (OR 0.60,95% CI 0.49-0.74) but not in term infants 5. This strengthens the case for PEEP use in preterm resuscitation specifically.
Bottom Line
While the Jackson-Rees circuit can technically deliver PEEP, its unreliability makes it a suboptimal choice for neonatal resuscitation. If this is your only available device, ensure operators receive specific training in tail-piece occlusion technique and understand that consistent PEEP delivery requires significant manual skill that most clinicians do not possess 2.