What causes the 'rabbit ears' pattern on the arterial pressure waveform during intra‑aortic balloon pump (IABP) therapy?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: April 18, 2026 • View editorial policy

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.

"Rabbit Ears" on Arterial Waveform with IABP

The "rabbit ears" pattern on arterial pressure waveforms during IABP therapy is caused by early balloon inflation, which creates two distinct systolic peaks—the first representing the patient's native systolic pressure and the second representing premature balloon inflation during late systole rather than early diastole.

Mechanism of the Malfunction

The rabbit ears appearance occurs due to timing error where the balloon inflates too early in the cardiac cycle. Instead of inflating at the dicrotic notch (aortic valve closure marking the beginning of diastole), the balloon inflates while the aortic valve is still open during late systole 1, 2.

What Creates the Two Peaks:

  • First peak: The patient's native left ventricular systolic ejection
  • Second peak: Premature balloon inflation creating a second pressure wave while blood is still being ejected

This creates the characteristic "double hump" or "rabbit ears" appearance on the arterial pressure tracing.

Hemodynamic Consequences

Early inflation is hemodynamically detrimental because:

  • Increases afterload against which the left ventricle must eject, directly opposing the intended therapeutic effect
  • Increases myocardial oxygen consumption rather than reducing it
  • Reduces stroke volume by creating resistance during systolic ejection
  • Negates the benefits of counterpulsation 3, 2

The IABP is designed to provide hemodynamic support through diastolic augmentation (increasing coronary perfusion) and systolic unloading (reducing afterload), but early inflation achieves the opposite effect 4.

Timing Correction Algorithm

Immediate action required:

  1. Identify the dicrotic notch on the arterial waveform—this marks aortic valve closure and the proper inflation point
  2. Delay inflation timing on the IABP console until the balloon inflates precisely at or just after the dicrotic notch
  3. Verify correct timing by confirming:
    • Single systolic peak (patient's native systole)
    • Sharp V-shaped dip after systole (balloon deflation)
    • Elevated diastolic plateau (balloon inflation/augmentation)
    • Reduced end-diastolic pressure before next systole

Critical Pitfall: Waveform Source Matters

Important caveat: Timing discrepancies exist based on which arterial pressure waveform you use for synchronization 1. Studies demonstrate:

  • IABP catheter lumen pressure: Delays of 74 ± 29 ms for both inflation and deflation landmarks
  • Radial artery pressure: Even greater delays of 175 ± 50 ms due to peripheral wave transmission

These delays can create conflicting indications of proper timing and alter IABP efficacy 1. When possible, use the most proximal arterial pressure source (central aortic pressure provides the most accurate timing reference), though this is rarely practical in clinical settings.

Verification of Proper Timing

Once corrected, the waveform should show:

  • Diastolic augmentation higher than the patient's systolic pressure
  • End-diastolic pressure reduction below the patient's unassisted end-diastolic pressure
  • Smooth, single systolic peak without the double-hump appearance
  • Clear separation between systolic ejection and diastolic augmentation phases 5, 6

The IABP operates on a counterpulsation principle—deflating just before ventricular ejection and inflating immediately after aortic valve closure 6, 2. Any deviation from this timing, particularly early inflation, produces the rabbit ears pattern and compromises patient outcomes.

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.