The Blue Vent is an Air Intake Port That Prevents Vacuum Occlusion
The blue piece making a suction noise is the air vent (also called a "pigtail" or sump port) on a double-lumen nasogastric tube, and the noise indicates it's functioning properly by allowing air to enter and break the vacuum seal that would otherwise cause gastric mucosa to occlude the drainage holes.
What's Happening
The suction noise you're hearing is actually a good sign - it means:
- Air is being drawn through the vent lumen while suction is applied to the main drainage lumen
- This air intake prevents the gastric wall from being sucked against and blocking the drainage holes
- The turbulent air flow through the smaller vent opening creates the characteristic "hissing" or "sucking" sound
Why This Design Exists
Double-lumen tubes (like Salem Sump) have two channels:
- Main drainage lumen (larger) - removes gastric contents
- Air vent lumen (smaller, with the blue connector) - continuously admits atmospheric air
Without the air vent, even moderate suction (as low as 60 mm Hg) causes persistent mucosal occlusion of all drainage holes, rendering the tube completely non-functional 1. Research demonstrates that at suction levels exceeding 90 mm Hg, 100% of single-lumen tubes malfunction due to mucosal blockage 1.
What You Should Do
Keep the blue vent open to air at all times - this is critical for proper function:
- Never cap, clamp, or instill anything through the blue vent
- Never connect the vent to suction
- The noise is normal and expected during active suctioning
- If the noise stops suddenly, check for vent occlusion (mucus, tape, etc.)
Common Pitfall to Avoid
Do not attempt to silence the vent by capping it - this defeats its entire purpose and will cause the tube to malfunction through mucosal occlusion. The supplementary small openings created by the vent system permit "turbulent evacuation characteristic of large-bore patency" even at high suction levels up to 300 mm Hg 1.
The noise indicates your NGT is working exactly as designed - the air vent is maintaining patency by preventing vacuum-induced mucosal blockage of the drainage ports.