High Volume, Low Pressure Suction Systems Are Required to Achieve –20 cm H₂O
When hospital wall suction cannot be reliably regulated, use a high volume, low pressure system such as a Vernon-Thompson pump or wall suction with a pressure-reducing adaptor to safely achieve –20 cm H₂O negative pressure. 1
Why System Type Matters
The critical distinction is between high volume, low pressure versus other suction configurations. The BTS guidelines explicitly state that high volume, low pressure systems (–10 to –20 cm H₂O) are recommended for chest drainage 1. This recommendation stems from physiologic principles and safety concerns:
Systems to AVOID:
High pressure, high volume systems - These generate excessive air flow that can cause:
- Air stealing
- Hypoxemia
- Perpetuation of persistent air leaks 1
High pressure, low volume systems - These should also be avoided 1
Recommended Solution:
Use a Vernon-Thompson pump or wall suction with an adaptor to reduce pressure 1. These systems provide:
- High volume capacity (15-20 L/min air flow volume) 1
- Controlled low pressure (–10 to –20 cm H₂O range)
- Ability to maintain consistent negative pressure despite variations in air leak
Physiologic Rationale
Normal intrapleural pressures range from –8 cm H₂O during inspiration to –3.4 cm H₂O during expiration 1. Because various physiological factors influence the actual suction applied to the pleural space, the system must have sufficient volume capacity to accommodate air leaks while maintaining the desired pressure range 1.
Critical Safety Considerations
- Never apply suction immediately after tube insertion - Wait 48 hours to avoid re-expansion pulmonary edema 1
- Patients requiring suction should only be managed on specialized lung units with experienced medical and nursing staff 1
- The pressure-reducing adaptor is essential when using wall suction to prevent excessive negative pressure
Common Pitfall
The most dangerous error is using unregulated high-pressure wall suction directly without an adaptor. This can generate dangerously high negative pressures that damage lung tissue and worsen air leaks. Always interpose a pressure-reducing device between wall suction and the drainage system.