Are Wave or Shore Chest Drainage Systems Like the Vernon-Thompson Pump?
Yes, if the Wave and Shore systems are designed as high-volume, low-pressure suction systems, they would function similarly to the Vernon-Thompson pump for chest drainage. The critical characteristic is not the brand name but whether these systems deliver the recommended specifications of high air flow volume (15-20 L/min) with low negative pressure (-10 to -20 cm H₂O) 1.
Understanding the Vernon-Thompson Pump Standard
The Vernon-Thompson pump is specifically cited in BTS guidelines as an example of the ideal chest drainage system characteristics 1:
- High volume capacity: Ability to handle 15-20 L/min air flow
- Low pressure: Maintains -10 to -20 cm H₂O suction
- Stable pressure maintenance: Does not generate excessive negative pressure that could cause complications
What Makes a System "Like" the Vernon-Thompson Pump
Any chest drainage system—whether called Wave, Shore, or another brand—would be functionally equivalent to a Vernon-Thompson pump if it meets these specifications 1:
Required Characteristics:
- High air flow volume capability (15-20 L/min minimum)
- Regulated low pressure (-10 to -20 cm H₂O)
- Pressure stability during air leaks
- Avoidance of high-pressure generation that risks air stealing, hypoxemia, or perpetuation of air leaks
Systems to Avoid
The guidelines explicitly warn against 1:
- High pressure, high volume systems: Can generate excessive air flow causing air stealing and hypoxemia
- High pressure, low volume systems: Inadequate for effective drainage
- Unregulated wall suction without pressure adaptors: May deliver unpredictable pressures
Modern System Considerations
Contemporary chest drainage systems fall into categories 2:
Dry Systems
- Generally maintain stable negative pressure
- Most (except Sentinel Seal) achieve 17-30 L/min air flow under normal conditions
- Better pressure maintenance during air leaks
- Safer for patient transport
Wet Systems
- Achieve 24-27 L/min air flow normally
- Can increase to 26-49 L/min with higher suction
- Caveat: Negative pressure becomes unstable due to water loss phenomenon
Single-Chamber Systems
- Achieve 22-28 L/min air flow
- Warning: Increase both air flow AND negative pressure with higher wall suction, potentially exceeding safe limits
Clinical Application Algorithm
To determine if Wave/Shore systems are appropriate:
Check manufacturer specifications for:
- Maximum air flow capacity (must be ≥15-20 L/min)
- Pressure regulation mechanism (must maintain -10 to -20 cm H₂O)
- Pressure stability during air leaks
If specifications match Vernon-Thompson characteristics → System is functionally equivalent and appropriate
If system lacks pressure regulation → Must add wall suction adaptor to reduce pressure 1
If system cannot achieve required air flow → Not suitable for cases with large air leaks
Important Caveats
Timing matters: Suction should not be applied immediately after chest tube insertion but added after 48 hours for persistent air leak or failure to re-expand 1. Early suction application, particularly in primary pneumothorax, risks re-expansion pulmonary edema 1.
Specialist setting required: Patients requiring suction should only be managed in lung units with specialist medical and nursing experience 1.
Pressure variability warning: Even well-designed systems can deliver variable intrapleural pressures if connecting tubes become obstructed with fluid 3. Complete patency of drainage tubing is essential for predictable pressure delivery.
Bottom Line
The Vernon-Thompson pump represents a functional specification rather than a specific device. Wave and Shore systems would be equivalent if they deliver high-volume, low-pressure suction within the recommended parameters. The key is verifying the technical specifications match the guideline requirements of 15-20 L/min air flow capacity with regulated -10 to -20 cm H₂O pressure 1.