Volume-Targeted Ventilation Mode: Settings and Considerations
In volume-targeted ventilation, set an initial tidal volume of 6 mL/kg predicted body weight with plateau pressure maintained ≤30 cmH₂O, as this approach reduces mortality in ARDS and prevents ventilator-induced lung injury across most clinical contexts. 1, 2, 3
Core Mechanism and Rationale
Volume-targeted ventilation operates by having the clinician set a specific tidal volume to be delivered, with the ventilator generating whatever pressure is necessary to achieve this volume within the set inspiratory time 4. This differs fundamentally from pressure-targeted modes where pressure is fixed and volume varies based on lung compliance.
The physiologic advantage is straightforward: controlling delivered volume prevents the overdistension and repetitive alveolar injury that drives ventilator-induced lung injury, particularly in diseased lungs with heterogeneous compliance 1.
Initial Ventilator Settings
Tidal Volume
- Start at 6 mL/kg predicted body weight (PBW) for ARDS and most acute respiratory failure 1, 2, 3
- PBW calculation: Males = 50 + 0.91[height(cm) - 152.4] kg; Females = 45.5 + 0.91[height(cm) - 152.4] kg 3
- Acceptable range: 4-8 mL/kg PBW, with lower volumes (even <6 mL/kg) potentially beneficial in severe ARDS 1
- For surgical patients: 6-8 mL/kg PBW initially 5
Plateau Pressure
- Maintain plateau pressure ≤30 cmH₂O as the upper safety limit 1, 2, 3
- This threshold prevents excessive alveolar stretch even when chest wall compliance is reduced
- If plateau pressure exceeds 30 cmH₂O despite appropriate tidal volume, consider permissive hypercapnia rather than increasing volume 4, 6
Driving Pressure
Driving pressure (ΔP = plateau pressure - PEEP) is the single best predictor of outcomes in ARDS, with values >15 cmH₂O indicating high risk 1. This reflects the functional lung size available for ventilation. When driving pressure is low despite standard tidal volumes, you may have more flexibility; when elevated, prioritize reducing it over strict adherence to other targets.
PEEP Settings
- Initial PEEP: 5 cmH₂O 5
- For moderate-to-severe ARDS: use higher PEEP levels (typically 10-15 cmH₂O) 2
- PEEP prevents end-expiratory alveolar collapse and reduces repetitive recruitment/derecruitment injury 4
- In obstructive disease, PEEP may offset intrinsic PEEP to reduce work of breathing, but setting PEEP above intrinsic PEEP is harmful 4
Respiratory Rate and Timing
- Rate: 10-15 breaths/min for restrictive disease; adjust based on minute ventilation needs 4
- Inspiratory time (Ti): Must be sufficient to deliver the set volume without excessive peak pressures
- I:E ratio: 1:2 to 1:1.4 depending on pathology 4
Disease-Specific Adjustments
ARDS
- Tidal volume: 6 mL/kg PBW (strong evidence for mortality reduction) 1, 2, 3
- Plateau pressure: ≤30 cmH₂O
- Higher PEEP (10-15 cmH₂O) combined with low tidal volume shows mortality benefit 1
- Permissive hypercapnia is acceptable with pH >7.20 4, 6
- Consider prone positioning if PaO₂/FiO₂ <150 2
COPD/Obstructive Disease
- Allow longer expiratory time to prevent gas trapping 4
- Lower I:E ratio (more time for exhalation)
- Monitor for intrinsic PEEP; set external PEEP cautiously 4
- Target higher pCO₂ if baseline is chronically elevated (inferred from high bicarbonate) 4
Neuromuscular Disease/Chest Wall Deformity
- May require higher pressures (10-15 cmH₂O) to overcome chest wall stiffness 4
- Tidal volume: 6 mL/kg PBW
- Higher respiratory rates (15-25 breaths/min) may be needed 4
- I:E ratio closer to 1:1 4
Critical Pitfalls to Avoid
Do not chase normal blood gases aggressively: Permissive hypercapnia with pH >7.20 is safer than high tidal volumes or pressures 4, 6
Plateau pressure >30 cmH₂O is a hard stop: If this occurs with appropriate tidal volumes, accept higher CO₂ rather than increasing volume 1, 4
Volume-targeted modes do NOT compensate for leaks like pressure modes do: This is a major disadvantage in non-invasive ventilation where mask leak is inevitable 4. For NIV, pressure-targeted modes are strongly preferred 4
Driving pressure matters more than tidal volume alone: A patient receiving 6 mL/kg with driving pressure of 18 cmH₂O is at higher risk than one receiving 7 mL/kg with driving pressure of 12 cmH₂O 1
Avoid sudden pressure increases: Volume modes deliver whatever pressure is needed, which can cause discomfort and patient-ventilator dyssynchrony if lung compliance suddenly worsens 4
Monitoring and Adjustment
- Measure plateau pressure regularly (at least every 4-6 hours and with any clinical change)
- Calculate driving pressure with each plateau pressure measurement
- Adjust tidal volume downward if plateau pressure approaches 30 cmH₂O or driving pressure exceeds 15 cmH₂O
- Monitor for patient-ventilator dyssynchrony: Volume modes are less forgiving of changes in patient effort
- Target SpO₂ 92-97% or PaO₂ 70-90 mmHg; avoid hyperoxia 1
When Volume-Targeted Ventilation Is NOT Preferred
Volume-targeted modes are rarely used for non-invasive ventilation in the UK and most centers because they cannot compensate for mask leak, deliver uncomfortable pressure spikes, and lack the expiratory pressure support needed for CO₂ clearance 4. For NIV, pressure-targeted bi-level modes are strongly preferred.
Evidence Strength
The recommendation for low tidal volume (6 mL/kg PBW) with plateau pressure ≤30 cmH₂O in ARDS represents Level 1++ evidence with strong recommendations from multiple high-quality guidelines 1, 2, 3. The landmark ARDS Network trial demonstrated absolute mortality reduction from 40% to 31% with this strategy, making it one of the most robust findings in critical care medicine.