Adjusting Expiratory Time and Trigger Settings in SIMV for Poor Synchrony
Shorten the expiratory time setting immediately—on some ventilators, setting a long expiratory time creates a long "lock out" period that prevents patient-triggered breaths, directly causing the poor synchrony and reduced tolerance you're observing. 1
Understanding the Problem
The issue stems from how SIMV (Synchronized Intermittent Mandatory Ventilation) handles the relationship between expiratory time and patient triggering:
- In assist/control modes like SIMV, the ventilator uses a "lock out" period to prevent breath stacking 1
- As respiratory rate increases, the lock out period must shorten proportionally 1
- Critical pitfall: On some ventilators, setting a long expiratory time automatically sets a long lock out period, which blocks patient-triggered breaths and causes intolerance 1
Specific Adjustment Algorithm
Step 1: Adjust Expiratory Time Based on Respiratory Rate
Target I:E ratio of at least 1:2, with longer expiration for obstructive disease:
- At respiratory rate 15 breaths/min: Use inspiratory time 1.2-1.6 seconds (30-40% of cycle time) 2
- For concurrent obstructive airways disease: Use shorter inspiratory time (~30% IPAP time) to allow adequate exhalation, especially at higher respiratory rates 2
- For restrictive disease: Use longer inspiratory time (~40% IPAP time) 2
Step 2: Optimize Trigger Sensitivity
Address intrinsic PEEP (PEEPi) which is the primary cause of trigger failure:
- Add EPAP (expiratory positive airway pressure) 3-5 cm H₂O to offset intrinsic PEEP 1
- This helps triggering by reducing the inspiratory effort needed before airflow begins 1
- Although PEEPi may reach 10-15 cm H₂O in severe COPD, EPAP levels >5 cm H₂O are rarely tolerated 1
Step 3: Consider Alternative Mode if Synchrony Remains Poor
If adjustments fail, switch to timed/ACV mode rather than continuing to struggle with patient-triggered breaths:
- Set mandatory breaths to match the patient's natural ventilatory pattern 1
- This is particularly important when the patient has advanced respiratory failure or insufficient respiratory effort to reliably trigger breaths 1
Key Mechanisms of Asynchrony
Patient-ventilator asynchrony results from:
- Undetected inspiratory effort
- Delay in response to start of inspiration
- Delay in detection of end of breath
- Excessive leakage 1
These causes may be difficult to resolve individually—hence the recommendation to use timed mode as an alternative 1
Common Pitfalls to Avoid
Don't assume all ventilators behave the same—the relationship between expiratory time and lock out period varies by manufacturer 1
Don't ignore the role of EPAP—it serves multiple functions: eliminates exhaled CO₂, aids triggering, and counteracts upper airway collapse 1
Don't persist with spontaneous triggering in advanced respiratory failure—patients may cease making spontaneous effort when "captured" or have hypoxic drive dependence 1
Don't overlook excessive leakage—this impairs both inspiratory and expiratory trigger function 1