Initial Pressure Support Ventilation Settings for Adults
Start with a pressure support level of 8 cmH₂O, PEEP of 5 cmH₂O, FiO₂ titrated to SpO₂ 88-92% (or >92% in asthma), flow trigger sensitivity of 0.7-2.0 L/min, and default flow-cycle at 25% of peak inspiratory flow.
Pressure Support Level
Begin with 8 cmH₂O of pressure support (the difference between IPAP and EPAP) 1. This provides adequate ventilatory assistance while allowing patient adaptation to positive pressure. The minimum recommended pressure support is 4 cmH₂O, but 8 cmH₂O offers more robust initial support 1.
- IPAP (Inspiratory Positive Airway Pressure): Start at 13 cmH₂O (8 cmH₂O support + 5 cmH₂O PEEP)
- EPAP (Expiratory Positive Airway Pressure)/PEEP: Start at 5 cmH₂O
- Adjust pressure support upward in 1-2 cmH₂O increments if tidal volumes are inadequate or work of breathing remains excessive 1
- Maximum pressure support should not exceed 20 cmH₂O 1
PEEP Settings
Initiate PEEP at 5 cmH₂O 1, 2. This baseline level prevents alveolar collapse at end-expiration while minimizing hemodynamic compromise. The evidence shows:
- PEEP of 5 cmH₂O is standard across multiple guidelines for initial settings 2, 3
- In ARDS or severe hypoxemia, higher PEEP (>5 cmH₂O) is indicated, but for general pressure support initiation, 5 cmH₂O is appropriate 4
- Avoid zero PEEP, which increases inspiratory trigger delay and patient-ventilator asynchrony, particularly in COPD patients 5
FiO₂ Titration
Set initial FiO₂ based on clinical context:
- Target SpO₂ 88-92% for most patients, particularly those with COPD or chronic hypercapnia 2
- Target SpO₂ >92-96% for asthma patients 2
- Start with FiO₂ sufficient to achieve target, then titrate down as tolerated
- Most severely hypoxemic patients receive 100% FiO₂ initially, but this should be rapidly weaned once oxygenation stabilizes 3
Trigger Sensitivity
Use flow triggering at 0.7-2.0 L/min 6. Flow triggering is preferred over pressure triggering in modern practice:
- Flow trigger sensitivity of 0.7-2.0 L/min provides responsive triggering without auto-triggering 6
- Pressure triggering at -1 cmH₂O is an alternative but offers no significant advantage over flow triggering in patients without COPD 6
- More sensitive triggers (lower threshold) reduce inspiratory trigger delay and patient effort, but risk auto-triggering 7
- Less sensitive triggers increase work of breathing and may worsen patient-ventilator asynchrony 5, 7
Flow-Cycle (Expiratory Trigger)
Set flow-cycle at 25% of peak inspiratory flow (manufacturer default on most ventilators). This determines when the ventilator cycles from inspiration to expiration:
- Lower percentages (e.g., 10-15%) prolong inspiration, useful in restrictive disease
- Higher percentages (e.g., 40-50%) shorten inspiration, beneficial in obstructive disease to allow longer expiratory time
- Adjust based on patient-ventilator synchrony and I:E ratio requirements
Expected Tidal Volumes
With these settings, expect delivered tidal volumes of 6-8 mL/kg predicted body weight 8, 2. This aligns with lung-protective ventilation principles:
- Target visible chest rise without excessive volumes
- Monitor for adequate minute ventilation (typically 6-10 L/min)
- If tidal volumes are inadequate (<5 mL/kg), increase pressure support by 2 cmH₂O increments 1
Critical Adjustments Based on Pathology
For obstructive disease (COPD, asthma):
- Maintain PEEP at 5 cmH₂O (may need adjustment if significant auto-PEEP present)
- Ensure adequate expiratory time (I:E ratio 1:2 to 1:4)
- Target SpO₂ 88-92% 2
For restrictive disease or neuromuscular weakness:
- May require higher pressure support (10-15 cmH₂O) to achieve adequate tidal volumes
- I:E ratio closer to 1:1 or 1:2 2
- Target SpO₂ >92%
For severe hypoxemia/ARDS:
- Consider higher PEEP (8-10 cmH₂O initially, titrate upward as needed) 4
- Maintain lung-protective tidal volumes (6 mL/kg PBW)
- May require transition to volume control if pressure support inadequate
Common Pitfalls to Avoid
- Excessive pressure support: Leads to large tidal volumes, patient-ventilator dyssynchrony, and delayed weaning
- Zero PEEP: Increases work of breathing and trigger delay, particularly problematic in COPD 5
- Over-sensitive triggers: Causes auto-triggering and patient distress
- Inadequate monitoring: Reassess within 15-30 minutes and adjust based on blood gases, respiratory mechanics, and patient comfort