ASV Settings Conversion from PCV+
Start ASV with %MinVol at 165% (targeting your current minute ventilation of 10.8 L/min), PEEP 5 cmH₂O, and FiO₂ 30%, allowing the ventilator to automatically adjust pressure support and respiratory rate based on respiratory mechanics.
Initial ASV Settings
Based on your current PCV+ parameters, here's the direct conversion:
%MinVol setting: 165% (not the default 100%)
- Your current minute ventilation is 10.8 L/min
- Research demonstrates that the manufacturer's recommended 100%MinVol setting frequently fails to reduce work of breathing in patients with respiratory failure 1
- The optimal %MinVol setting that significantly reduces work of breathing averages 165% and can be identified when 1-3 mandatory breaths per minute begin to appear 1
- In 82% of patients, the 100%MinVol setting resulted in actual minute ventilation exceeding target, with inadequate pressure support 1
PEEP: 5 cmH₂O (maintain current setting)
- Transfer your current EPAP directly 2
FiO₂: 30% (maintain current setting)
Pressure limits:
How ASV Will Adjust Your Current Settings
ASV will automatically select tidal volume and respiratory rate based on respiratory mechanics 4, 5:
Expected tidal volume: 6-7 mL/kg IBW (ASV targets lung-protective ventilation)
Expected respiratory rate: Will likely decrease from your current rate of 12
- ASV typically selects lower rates when delivering adequate minute ventilation 4
- The ventilator will automatically adjust based on the minimum work of breathing principle
Critical Adjustment Algorithm
After initiating ASV at 165%MinVol:
Monitor for 5-10 minutes and observe:
If patient appears under-supported (increased work of breathing, respiratory distress):
- Increase %MinVol by 20% increments 1
- This increases pressure support and reduces patient effort
If excessive mandatory breaths occur (>3-5 per minute):
- Decrease %MinVol by 10% increments
- Patient may be over-sedated or has improved respiratory mechanics
Special consideration for COPD:
Key Monitoring Parameters
- Driving pressure: Should remain <15 cmH₂O (predictor of outcomes in ARDS) 7
- Plateau pressure: Should stay <30 cmH₂O 6
- Delivered tidal volume: 4-8 mL/kg IBW 6
- Arterial blood gases: Maintain similar to current PCV+ targets
- Patient-ventilator synchrony: ASV adjusts breath-by-breath to minimize dyssynchrony
Common Pitfall to Avoid
Do not start at 100%MinVol as recommended in the operator's manual - this setting was associated with inadequate pressure support and failed to reduce work of breathing in 82% of patients with respiratory failure 1. Starting at 165%MinVol provides the pressure support equivalent to your current PCV+ settings while allowing ASV's closed-loop algorithm to optimize the breathing pattern based on actual respiratory mechanics 8, 5.