Metallic vs Drug-Eluting Stents: Impact on FDG-PET Interpretation
Both metallic and drug-eluting stents can cause CT attenuation correction artifacts on PET images, but the key to accurate interpretation is reviewing BOTH non-attenuation-corrected (NAC) and attenuation-corrected (AC) images together—if uptake persists on NAC images, it represents true pathology rather than artifact. 1, 2
Understanding the Artifact Problem
The type of stent material (metallic bare-metal vs drug-eluting) is less important than recognizing that any dense metallic object causes overcorrection artifacts in CT-based attenuation correction, creating false "hot spots" on AC PET images that mimic infection 3. This occurs because:
- The CT-based attenuation correction algorithm overcompensates for the high-density metal
- Artifactual FDG uptake appears surrounding the metallic object on AC images only
- True pathologic uptake will be visible on BOTH AC and NAC images
Critical Interpretation Algorithm
When evaluating coronary stent uptake on FDG-PET/CT:
First, check NAC images: If increased FDG uptake is present on AC images adjacent to the stent, immediately review the NAC images 1, 2
Compare uptake patterns:
Assess uptake characteristics on NAC images:
- Location relative to stent (focal at stent site vs diffuse)
- Intensity compared to adjacent tissue
- Pattern (focal, linear along stent, or diffuse) 2
Correlate with CT findings: Look for supporting anatomical evidence such as:
- Soft tissue mass around stent
- Pericardial fluid collection
- Abscess formation 5
Drug-Eluting Stent Considerations
While drug-eluting stents contain metal and cause the same attenuation artifacts as bare-metal stents, they have additional interpretive considerations:
- Drug-eluting stents can cause local inflammatory reactions from the polymer coating or drug itself, potentially showing FDG uptake even without infection 6
- This inflammation-associated uptake may represent sterile inflammation rather than infection
- Clinical context (timing post-procedure, symptoms, inflammatory markers) becomes crucial for differentiation
Practical Pitfalls to Avoid
Common error: Interpreting any uptake near a stent on AC images as infection without checking NAC images—this leads to false positives 3
Modern scanners: Newer PET/CT devices with improved metal artifact reduction algorithms produce fewer artifacts, but NAC review remains essential when doubt exists 1
SUV measurements: Do not rely on SUV values or target-to-background ratios to differentiate infection from artifact or sterile inflammation—these are not validated for this purpose 2
When Stent Infection is Confirmed
If true FDG uptake is confirmed on NAC images with supporting clinical/anatomical findings 5, 7:
- Consider ECG-gated acquisition for better anatomical localization 5
- Evaluate for complications: in-stent restenosis, pseudoaneurysm, pericardial involvement 7
- Assess whole body for septic emboli and infection source 2
The distinction between metallic and drug-eluting stents matters less for artifact generation than for understanding potential causes of true uptake—both require the same rigorous NAC/AC comparison for accurate interpretation.