Measuring Aortic Stenosis and Aortic Regurgitation by TTE
For aortic stenosis, measure three primary parameters: peak jet velocity (≥4 m/s = severe), mean gradient (≥40 mmHg = severe), and aortic valve area by continuity equation (<1.0 cm² = severe); for aortic regurgitation, use vena contracta width (<3 mm = mild, >6 mm = severe) and PISA method as primary quantitative measures, supplemented by additional parameters including pressure half-time and diastolic flow reversal in the descending aorta. 122
Aortic Stenosis Measurement
Three Primary Hemodynamic Parameters
The assessment requires integration of all three parameters, as each provides complementary information 1:
1. Peak Jet Velocity
- Obtain continuous wave (CW) Doppler from multiple acoustic windows (apical, right parasternal, suprasternal)
- Use a dedicated small dual-crystal CW transducer for optimal signal
- Severe AS: ≥4 m/s
- Common pitfall: Misalignment of the beam—this is the most frequent source of gradient underestimation
2. Mean Transvalvular Pressure Gradient
- Calculate by averaging instantaneous gradients over the ejection period
- Cannot be calculated from mean velocity alone—must average instantaneous mean gradients
- Severe AS: ≥40 mmHg
3. Aortic Valve Area (AVA) by Continuity Equation
- Severe AS: <1.0 cm²
- This measures effective orifice area, which is the primary predictor of clinical outcome 1
Continuity Equation Components
LVOT Diameter Measurement:
- Measure in parasternal long-axis view
- Inner edge to inner edge (septal endocardium to anterior mitral leaflet)
- Measure in mid-systole
- Critical pitfall: LVOT diameter is the most common source of error—it is squared in the equation, so small measurement errors are magnified 1
LVOT Velocity:
- Use pulsed wave Doppler from apical approach
- Position: Apical five-chamber view or apical long-axis view
- Place sample volume just proximal to aortic valve (0.5-1.0 cm below if flow acceleration occurs at annulus)
- Obtain smooth velocity curve without spectral dispersion 1
Aortic Valve Velocity:
- Use CW Doppler from multiple windows
- Record highest velocity obtained
Serial Measurements
When following patients over time 1:
- Always record from the same window with same quality
- Report which window yielded highest velocities
- LVOT diameter rarely changes in stable adults—consider using the same measurement to reduce variability
- If AVA changes significantly, verify all equation components
Aortic Regurgitation Measurement
Primary Quantitative Methods
The color flow jet area alone is NOT recommended for quantification 22. Use these validated methods:
1. Vena Contracta (VC) Width (Preferred Semi-Quantitative Method)
- Measure in parasternal long-axis view (apical if unavailable)
- Nyquist limit: 50-60 cm/s
- Measure the smallest flow diameter immediately below the aortic valve in the LVOT
- Mild AR: <3 mm
- Severe AR: >6 mm
- Represents an estimate of effective regurgitant orifice area (EROA) 22
Technical approach:
- Optimize color gain/scale
- Identify three jet components: flow convergence zone, vena contracta, jet turbulence
- Reduce color sector size to maximize frame rate
- Use zoom function
- Use cine-loop to find best frame
- Measure perpendicular to jet direction 2
2. PISA (Proximal Isovelocity Surface Area) Method (Quantitative)
- Provides EROA and regurgitant volume
- Downward shift zero baseline to obtain hemispheric PISA
- Measure PISA radius at first aliasing
- Use CW Doppler of AR jet to calculate EROA and regurgitant volume 22
Jet Width Ratio (Semi-Quantitative)
- Measure maximum color jet diameter in diastole immediately below aortic valve
- Normalize to LVOT diameter (parasternal long-axis view)
- Jet width ratio >65% strongly suggests severe AR 2
- Limitation: Assumes circular orifice—less accurate with bicuspid valves or irregular orifices
Additional Supporting Parameters
CW Doppler Pressure Half-Time:
- Obtained from apical five-chamber view
- Severe AR: <200 ms (rapid pressure equalization)
- Note: Affected by LV compliance and chronicity 2
Diastolic Flow Reversal in Descending Aorta:
- Use pulsed wave Doppler in proximal descending or abdominal aorta
- Holodiastolic flow reversal with end-diastolic velocity >20 cm/s suggests severe AR 2
2D/M-Mode Findings:
- Flail valve is specific for severe AR
- High-frequency fluttering of anterior mitral leaflet
- Premature mitral valve closure (sign of acute severe AR with elevated LV end-diastolic pressure)
- Reverse doming of anterior mitral leaflet 22
Left Ventricular Assessment
- LV diameters and volumes (preferably Simpson's biplane method)
- LV ejection fraction
- Index all measurements to body surface area
- LV dilatation is sensitive for chronic severe AR but absent in acute severe AR
Integration of Severity
Combine multiple parameters rather than relying on single measurement 22:
- Quantitative methods (VC, PISA) form the foundation
- Support with additional Doppler parameters
- Assess LV remodeling and function
- Consider chronicity (acute vs. chronic)
Common Pitfalls
Aortic Stenosis:
- Beam misalignment causing velocity underestimation—always use multiple windows
- LVOT diameter measurement errors—measure carefully in mid-systole, inner edge to inner edge
- Incorrect mean gradient calculation—must average instantaneous gradients
- Low-flow states with LV dysfunction may underestimate true stenosis severity 1
Aortic Regurgitation: