Aliquots in Clinical Laboratory Testing
Definition and Purpose
Aliquots are smaller, divided portions of a biological sample (such as blood, plasma, serum, or CSF) that are separated from a larger parent specimen and stored in individual containers for laboratory testing or biobanking. The primary purpose of aliquoting is to avoid repeated freeze-thaw cycles of the entire sample, which can degrade analytes and compromise test accuracy 1.
Preparation of Aliquots
General Principles
Aliquots should be prepared in low-protein-binding polypropylene tubes using polypropylene pipette tips, with screw-cap tubes containing rubber O-rings to prevent evaporation during storage 1.
Volume Considerations
- Small aliquots (generally ≤0.5 mL) are recommended to minimize unnecessary freeze-thaw cycles 1
- For plasma and serum, consider even smaller volumes (55 µL, 200 µL, or 500 µL) depending on anticipated testing needs 1
- Fill tubes to at least 75% of tube capacity to reduce variability caused by differences in sample volume-to-tube surface area ratios 1
CSF-Specific Aliquoting Protocol
After centrifugation at 2000 × g for 10 minutes at room temperature (20-25°C):
- Transfer CSF to a clean 30-mL polypropylene tube without disturbing the pellet
- Invert briefly to mix
- Divide into 0.5-mL aliquots in 1-mL polypropylene tubes 1
- Freeze at -80°C (flash freezing on dry ice is not necessary)
Blood Sample Aliquoting
For plasma and serum specimens 1:
- Use validated polypropylene tubes or straws
- Document the total volume obtained and individual aliquot volumes
- Minimize time between centrifugation and freezing
- Record hemolysis (H1-H6 scale, with H3+ being significant) and lipemia status
Storage Recommendations
Temperature Requirements
Long-term storage should be at -80°C or in liquid nitrogen 1. Short-term storage varies by specimen type:
- Fresh CSF for Aβ/tau testing: 2-8°C for up to 14 days, or room temperature (20-25°C) for up to 2 days 1
- Plasma aliquots: Demonstrated stability varies by analyte; some remain stable at 2-8°C for 96 hours, while others degrade within hours 2
Critical Storage Considerations
- Store samples upright to avoid interaction with tube caps, which may not be polypropylene 1
- If using dry ice for shipment, vent the headspace or allow samples to equilibrate at -80°C for 9 hours before thawing 1
- Use disposable thermometers to monitor temperature during transportation 1
Management and Documentation
Essential Documentation
Record the following for each aliquot 1:
- Type of secondary container (tube or straw)
- Time between centrifugation and freezing
- Storage temperature and location
- Number of freeze-thaw cycles (critical for quality assurance)
- Duration of storage
- Degree of hemolysis and lipemia
Quality Assurance Considerations
Repeated freeze-thaw cycles degrade samples, so the number should be recorded and standardized among samples in the same study to avoid bias 3. This is why creating multiple small aliquots upfront is preferable to repeatedly thawing a single large sample.
Common Pitfalls and Solutions
Analyte Stability Issues
Five common biochemical analytes show reduced stability in primary plasma tubes: aspartate aminotransferase (7.9 hours), glucose (2.7 hours), lactate dehydrogenase (2.9 hours), inorganic phosphate (6.2 hours), and potassium (4.7 hours) at 2-8°C 2. For these analytes, plasma aliquots provide better stability for add-on testing.
Sample Degradation Prevention
- Hemolysis: Mitigated by skillful phlebotomy, use of 21-gauge or larger needles, and proper tube handling 3
- RNA degradation: Prevented by using stabilizing reagents (PAXgene or Tempus tubes) 3
- Inconsistent volumes: Ensure accurate blood draw volumes and proper tube mixing 3
Tube Material Considerations
For Aβ42 and other protein biomarkers, tube material matters significantly due to adsorption to non-polypropylene surfaces 1. Polystyrene tubes should ONLY be used for samples sent to clinical laboratories for routine testing, never for biobanking or specialized biomarker analysis.
Statistical Considerations for Aliquot Volume
The aliquot size directly impacts the limit of detection and false-negative rate for low-concentration analytes 4. For example, to maintain a false-negative rate <0.01%, minimum cell counts of 2000,1000, and 500 are required for 0.5%, 1%, and 2% aliquots, respectively. This is particularly relevant for nucleic acid amplification assays where target concentrations approach the detection threshold.