Mechanical Ventilation: Settings and Modes Overview
Mechanical ventilation operates through two fundamental control strategies—volume-targeted and pressure-targeted—with modes defined by how the ventilator initiates, delivers, and cycles breaths in relation to patient effort. 1
Core Control Strategies
Volume-Targeted Ventilation
- You set: Tidal volume (typically 4-8 mL/kg predicted body weight in ARDS) and inspiratory time
- The ventilator generates: Whatever pressure is necessary to deliver that volume
- The resulting pressure depends on: Lung/chest wall compliance and airway resistance 1
Pressure-Targeted Ventilation (Preferred in Modern Practice)
- You set: Inspiratory pressure (IPAP) and expiratory pressure (EPAP)
- The patient receives: Volume determined by lung mechanics and inspiratory time
- Key advantages: Constant pressure delivery (more comfortable), compensates for air leaks, maintains positive expiratory pressure 1
In pressure-targeted ventilation, the difference between IPAP and EPAP equals the pressure support level—this is the actual ventilatory assistance provided. 1
Primary Ventilator Modes
Controlled Mechanical Ventilation (CMV)
- Full ventilatory support—no patient effort required
- You set: Tidal volume (or pressure), respiratory rate, inspiratory time
- Machine delivers every breath at predetermined intervals
- Used when patient has minimal/no respiratory drive 2
Assist/Control (A/C or ACV)
- Guarantees a minimum respiratory rate but allows patient triggering
- Machine delivers preset mandatory breaths if patient doesn't trigger
- When patient triggers, machine delivers identical full-support breath
- Includes "lock-out" period to prevent breath stacking
- Also called Spontaneous/Timed (S/T) mode on NIV machines 2
Critical distinction: In A/C, triggered breaths delay the next mandatory breath, creating synchronization between patient and machine (SIMV concept). 2
Pressure Support (PS) / Assisted Spontaneous Breathing (ASB)
- Patient controls everything—triggers both breath initiation AND termination
- Patient determines respiratory rate and timing
- You only set the pressure support level
- Major caveat: If patient stops breathing, no support occurs (though most modern ventilators include backup rate of 6-8 breaths/minute) 2
Continuous Positive Airway Pressure (CPAP)
- Not true ventilatory support—primarily corrects hypoxemia
- Maintains constant positive pressure throughout respiratory cycle
- Recruits collapsed alveoli (similar to PEEP effect)
- Offsets intrinsic PEEP in COPD, reducing work of breathing
- Important limitation: In hyperinflated patients, CPAP may worsen inspiratory muscle function by further increasing lung volume 2
Critical Settings to Understand
Key Pressure Parameters
- Plateau pressure: Should be <30 cm H₂O in ARDS to prevent barotrauma 3
- PEEP: Prevents alveolar collapse, aids oxygenation
- Driving pressure (ΔP): Plateau pressure minus PEEP—emerging as better predictor of ARDS outcomes than tidal volume or plateau pressure alone 3
Tidal Volume Targets
For ARDS patients, use 4-8 mL/kg predicted body weight (not actual weight) 3:
- Males: 50 + 0.91 × [height(cm) - 152.4] kg
- Females: 45.5 + 0.91 × [height(cm) - 152.4] kg
Larger tidal volume gradients (difference between low and traditional volumes) show greater mortality benefit in meta-regression analysis 3.
Common Pitfalls
Terminology confusion: Manufacturers use different names for identical modes—"S/T mode," "SIMV," and "IE mode" may all refer to assist/control ventilation 2. Focus on understanding the physiological principle (does patient trigger? does machine guarantee rate?) rather than memorizing brand names.
CPAP flow requirements: In acute respiratory failure with high minute ventilation, peak inspiratory flows may exceed 60 L/min. Low-flow CPAP generators (adequate for sleep apnea) will fail to maintain pressure in distressed patients 2.
Lock-out periods in A/C mode: Setting long expiratory times creates long lock-out periods on some ventilators, causing poor patient tolerance and potential dyssynchrony 2.
Evidence-Based Approach
The 2016 BTS/ICS guidelines emphasize that pressure-targeted ventilation is preferred in UK practice (volume ventilators rarely used outside specialist centers) due to better patient comfort, leak compensation, and maintenance of expiratory positive pressure 1.
For ARDS specifically, the 2017 ATS/ERS/SCCM guidelines provide moderate-confidence evidence that low tidal volume strategies reduce mortality when combined with appropriate PEEP (RR 0.80,95% CI 0.66-0.98), with greater benefit when tidal volume gradient between intervention and control is larger 3.