RVOT Diameter and VTI Measurement Technique
Measure the RVOT diameter as a linear transversal dimension just proximal to the pulmonary valve at end-diastole from the parasternal short-axis or long-axis view, and obtain the RVOT VTI using pulsed-wave Doppler positioned at the same location where you measured the diameter. 1
RVOT Diameter Measurement
The distal RVOT diameter is the preferred measurement location according to ASE/EACVI guidelines 1:
- View: Parasternal long-axis or parasternal short-axis view
- Timing: End-diastole
- Location: Linear transversal dimension measured just proximal to the pulmonary valve
- Technique: Inner edge to inner edge measurement
There is also a proximal RVOT diameter option, measured from the anterior RV wall to the interventricular septal-aortic junction (parasternal long-axis) or to the aortic valve (parasternal short-axis), but this is less reproducible than the distal measurement 1.
Key Technical Points for Diameter:
- The distal RVOT measurement is more reproducible than the proximal measurement 1
- Ensure the view is not obliquely oriented with respect to the RVOT to avoid under/overestimation 1
- Document which window you used, as measurements are dependent on probe rotation and view 1
- Be aware that endocardial definition of the RV anterior wall is often suboptimal 1
RVOT VTI Measurement
Pulsed-wave Doppler technique 2, 3:
- Sample volume position: Place at the same level where you measured the RVOT diameter (just proximal to the pulmonary valve)
- View options:
- Technique: Trace the velocity envelope throughout systole to obtain the VTI in centimeters
Alternative Subcostal Approach:
Recent evidence demonstrates that RVOT VTI can be reliably obtained from a modified subcostal window when apical or parasternal views are suboptimal 2, 3. This is particularly useful in:
- Mechanically ventilated patients 3
- Critically ill patients with limited acoustic windows 2
- The modified subcostal approach allows RVOT VTI measurement in 100% of patients versus 81% with traditional parasternal views 3
Clinical Caveats
Important limitations to recognize 1:
- RVOT dimensions can be inaccurate in patients with chest and spine deformities
- This is a regional measure and may not reflect global RV size
- Limited normative data exists compared to left-sided measurements
- Risk of measurement error if the imaging plane is not properly aligned with the RVOT
Normal reference values 4:
- Normal RVOT VTI should be >13 cm when TAPSE >2 cm
- RVOT VTI >15 cm is expected when PASP <35 mmHg or PVR <1.6 Wood units
Practical Application
The RVOT diameter and VTI are used together to calculate RV stroke volume using the continuity equation 5:
RV Stroke Volume = RVOT cross-sectional area × RVOT VTI
Where RVOT cross-sectional area = π × (RVOT diameter/2)²
While RVOT and LVOT stroke volumes show moderate correlation (ICC 0.63-0.89), they cannot be used interchangeably due to significant bias 5. The measurements are repeatable and reliable when performed systematically, with repeatability coefficients of 2-3 mm for diameter and 2.1-2.8 cm for VTI 5.