Volume Assist-Control Mode Does Not Require Pressure Control or Pressure Support Settings
In volume assist-control (AC) mode, you do not set pressure control or pressure support—these are fundamentally different ventilation modes that are mutually exclusive. Volume AC is a volume-targeted mode where you set the tidal volume and inspiratory time, and the ventilator generates whatever pressure is necessary to deliver that volume 1.
Understanding the Fundamental Difference
Volume assist-control operates on a completely different principle than pressure-targeted modes:
Volume AC: You set the tidal volume to be delivered and the duration of inspiration (Ti). The ventilator automatically generates whatever pressure is needed to achieve this volume within the specified time 1.
Pressure Control/Pressure Support: You set the inspiratory pressure level. The resulting tidal volume becomes a function of lung/chest wall mechanics and inspiratory time—not a preset value 1.
These are distinct ventilation strategies—you cannot "set" pressure control or pressure support within volume AC mode because they represent different control variables.
What You Actually Set in Volume AC
When using volume assist-control, your primary settings are:
- Tidal volume (typically 6-8 mL/kg ideal body weight)
- Respiratory rate (backup rate)
- Inspiratory time or inspiratory flow pattern
- PEEP
- FiO2
- Trigger sensitivity
The ventilator will deliver the set tidal volume with each triggered or mandatory breath, generating variable pressures based on respiratory system mechanics 2, 3.
Common Pitfall: Confusing Mode Terminology
The confusion likely stems from ventilators offering multiple modes. Modern ventilators may have:
- Volume AC (volume-targeted)
- Pressure AC (pressure-targeted with backup rate)
- Pressure Support (pressure-targeted, patient-triggered only)
These are separate mode selections—not parameters you adjust within volume AC 1. If you want pressure-targeted ventilation, you must switch to a different mode entirely.
Clinical Considerations
Volume AC has specific characteristics compared to pressure modes 1:
- Delivers constant tidal volume regardless of changing lung mechanics
- May generate uncomfortable pressure spikes if flow doesn't match patient demand
- Allows immediate assessment of respiratory system mechanics through pressure monitoring
- Does not automatically compensate for circuit leaks (unlike pressure modes)
If patient-ventilator dyssynchrony occurs in volume AC due to fixed flow patterns, the solution is adjusting the inspiratory flow rate or flow waveform (square vs. decelerating), not adding pressure support 2, 3.