As an urgent‑care physician, can I determine whether an intra‑ocular stent (e.g., iStent, Hydrus) has become displaced?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: November 27, 2026 • View editorial policy

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.

Can You Determine if an Intraocular Stent Has Moved in Urgent Care?

No, you cannot reliably determine whether an intraocular stent (iStent or Hydrus) has become displaced in an urgent care setting—this requires specialized ophthalmologic examination with gonioscopy and/or anterior segment optical coherence tomography (AS-OCT), and the patient needs immediate referral to an ophthalmologist.

Why This Assessment Requires Specialized Equipment

The detection of stent malposition is technically challenging even for ophthalmologists:

  • Gonioscopy underestimates malposition: Standard gonioscopic examination misses a significant proportion of malpositioned devices. Research shows that 92% of iStent inject devices were visible on AS-OCT versus only 88% on gonioscopy, and 45.7% of devices were completely burrowed within the trabeculum and not visible on standard examination 1.

  • AS-OCT is the gold standard: Advanced imaging with anterior segment OCT is necessary to definitively resolve stent position and orientation 2. This technology is not available in urgent care settings.

  • Subtle malposition causes serious complications: Even subtle malposition that appears normal on casual examination can cause uveitis-glaucoma-hyphema (UGH) syndrome requiring surgical explantation 3.

Clinical Red Flags That Warrant Immediate Ophthalmology Referral

While you cannot definitively assess stent position, recognize these warning signs of potential malposition:

Device-related complications occur in 3-18% of cases 4:

  • Recurrent hyphema (blood in anterior chamber) - occurs in 0.5-36% depending on device type
  • Persistent or recurrent anterior uveitis (eye redness, pain, photophobia)
  • IOP spikes - elevated intraocular pressure (1-33% incidence)
  • Corneal edema - cloudy cornea (1.4-28% with Hydrus)
  • Uncontrolled glaucoma despite the stent

Specific Patterns by Device Type

iStent complications 4:

  • Malposition: 3-18% of cases
  • Obstruction: 1-13.2%
  • Most common adverse event in recent surveillance: malposition (37.3% of device problems) 5

Hydrus Microstent complications 4:

  • Malposition: 1.1% (though likely underreported)
  • Device malposition most common issue (34.9% of adverse events) 5
  • Can cause UGH syndrome even with subtle malposition 3
  • Late removal (>3 months) is more challenging as device becomes embedded in iris 6

Your Role in Urgent Care

Your critical function is recognition and appropriate referral:

  1. Take a focused history: When was the stent placed? Any recent trauma or eye manipulation? New symptoms since placement?

  2. Document presenting symptoms: Pain, redness, vision changes, photophobia, seeing blood in the eye

  3. Perform basic examination: Visual acuity, external examination for injection/inflammation, check for visible hyphema with penlight

  4. Do NOT attempt gonioscopy if you lack specialized training—this requires specific equipment and expertise

  5. Refer urgently to ophthalmology (same day or next day) for any concerning symptoms

Common Pitfall to Avoid

Do not reassure the patient that "everything looks fine" based on external examination alone. Malposition causing serious complications may not be visible without specialized angle examination. Recent data shows stent malposition is the most commonly reported adverse event (32.9% of all complications) 5, yet many cases are not visible on routine examination.

Early recognition and referral is critical because late explantation (>3 months) becomes significantly more difficult as devices become embedded in ocular tissues 6.

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.