What method is used to measure electrolytes in most clinical laboratories?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: August 23, 2025View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

Methods Used to Measure Electrolytes in Clinical Laboratories

Ion-selective electrode (ISE) technology is the predominant method used to measure electrolytes in most clinical laboratories. 1 This technology has largely replaced older methods such as flame photometry, atomic absorption spectrophotometry, and coulometry for the determination of key electrolytes including potassium, sodium, calcium, and chloride.

Types of Ion-Selective Electrode Methods

There are two main approaches to ISE technology used in clinical laboratories:

  1. Direct ISE Method

    • Measures the activity of ions directly in undiluted samples
    • Used primarily in blood gas analyzers and point-of-care devices
    • Provides more accurate measurements in samples with abnormal protein concentrations
    • Not affected by sample dilution
  2. Indirect ISE Method

    • Involves dilution of the sample before measurement
    • Most commonly used in high-throughput central laboratory analyzers
    • Can be affected by abnormal protein concentrations, leading to "pseudohyper" or "pseudohypo" electrolyte values 2

Electrolytes Commonly Measured

ISE technology is routinely used to measure:

  • Sodium (Na+)
  • Potassium (K+)
  • Chloride (Cl-)
  • Calcium (Ca2+)
  • Magnesium (Mg2+)
  • Bicarbonate (HCO3-)

Technical Aspects and Considerations

  • ISE analyzers use specialized membranes that are selective for specific ions
  • Six different types of ion-selective sensors exist: glass membrane, solid phase, fluid membrane, carrier, gas-sensitive, and enzyme electrodes 1
  • Modern analyzers can process multiple electrolytes simultaneously from a single sample
  • Automated systems have significantly improved throughput and standardization
  • Typical coefficient of variation for electrolyte measurements ranges from 0.49%-2.23% for within-run precision 3

Clinical Implications of ISE Technology

Advantages

  • Rapid analysis time (results available in minutes)
  • Requires small sample volumes
  • High accuracy and precision when properly calibrated
  • Convenience in operation 4

Limitations and Considerations

  • Protein concentration effects: In patients with abnormal protein levels (particularly hypoproteinemia), indirect ISE methods may produce inaccurate results
  • For critically ill patients with abnormal protein concentrations, direct ISE methods provide more accurate measurements for clinical decision-making 2
  • Regular calibration and quality control are essential for reliable results

Special Situations

In specific clinical scenarios such as:

  • Critically ill patients
  • Patients with significant protein abnormalities
  • Patients requiring urgent electrolyte assessment

Direct ISE methods (such as those used in blood gas analyzers) may be preferred over indirect ISE methods to avoid measurement errors related to protein concentration effects 2.

Quality Control

For optimal clinical utility, ISE-based electrolyte analyzers require:

  • Careful and accurate calibration
  • Appropriate sample preparation
  • Adjustment of measuring conditions to sample characteristics
  • Consistent internal and external quality control 1

ISE technology represents a significant advancement in clinical chemistry and remains the standard method for electrolyte measurement in modern clinical laboratories due to its speed, accuracy, and reliability when properly implemented and maintained.

References

Research

Ion-selective potentiometry in clinical chemistry. A review.

Medical progress through technology, 1988

Research

Effects of total plasma protein concentration on plasma sodium, potassium and chloride measurements by an indirect ion selective electrode measuring system.

Critical care and resuscitation : journal of the Australasian Academy of Critical Care Medicine, 2005

Research

[The development of model MI-921A automatic clinical electrotyte analyzer].

Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation, 1997

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.