Switching to BPAP or BPAP ST During PAP Therapy Desensitization
Yes, it is reasonable and recommended to switch to BPAP or BPAP ST mode during PAP therapy desensitization if a patient is experiencing discomfort with CPAP. 1, 2
Indications for Switching to BPAP
- Patient discomfort with CPAP: The American Academy of Sleep Medicine explicitly states that "if the patient is uncomfortable or intolerant of high pressures on CPAP, the patient may be tried on BPAP" 1
- Continued obstructive events: If there are continued obstructive respiratory events at 15 cm H₂O of CPAP during titration 1
- Pressure intolerance: BPAP should be considered for patients who are nonadherent to CPAP therapy because of pressure intolerance 3
- Bloating symptoms: Patients experiencing persistent bloating on CPAP may benefit from switching to BPAP 2
When to Consider BPAP ST Mode
- Treatment-emergent central apneas: A decrease in IPAP or setting BPAP in spontaneous-timed (ST) mode with backup rate may be helpful if treatment-emergent central apneas (complex sleep apnea) are observed during titration 1
- Specific patient populations: A backup rate (ST mode) should be used in all patients with:
- Central hypoventilation
- Significant number of central apneas
- Inappropriately low respiratory rate
- Unreliable triggering of IPAP/EPAP cycles due to muscle weakness 1
Initial BPAP Settings During Desensitization
- Starting IPAP: 8 cm H₂O 1, 2
- Starting EPAP: 4 cm H₂O 1, 2
- IPAP-EPAP differential: Maintain minimum of 4 cm H₂O and maximum of 10 cm H₂O 1
- Pressure adjustments: Increase IPAP and/or EPAP by at least 1 cm H₂O with intervals no shorter than 5 minutes 1
BPAP ST Settings (When Needed)
- Starting backup rate: Equal to or slightly less than the spontaneous sleeping respiratory rate (minimum of 10 breaths per minute) 1
- Adjustments: Increase backup rate in 1-2 breaths per minute increments every 10 minutes if goals not attained 1
- Inspiratory time: Set between 30-40% of the cycle time (60/respiratory rate) 1
Improving Patient Comfort During Desensitization
- Pressure reduction for comfort: If the patient awakens and complains that pressure is too high, restart at a lower pressure that the patient reports is comfortable enough to allow return to sleep 1
- Device parameters: Adjust pressure relief, rise time, and maximum/minimum IPAP durations to optimize patient comfort and synchrony 1
- Mask considerations: Perform mask refit, adjustment, or change in mask type whenever significant unintentional leak is observed or patient complains of mask discomfort 1
- Humidification: Add heated humidification if patient complains of dryness or significant nasal congestion 1
Clinical Benefits of Switching to BPAP
- Improved adherence: Patients who fail CPAP therapy due to pressure intolerance can achieve better adherence with BPAP (7.0 vs 2.5 hours/night) 4
- Better symptom control: BPAP can provide better control of sleep-disordered breathing symptoms compared to failed CPAP therapy 4
- Lower expiratory pressures: BPAP allows for lower expiratory pressures compared to equivalent CPAP (10 vs 16.8 cm H₂O) 4
Monitoring and Follow-up
- Regular monitoring: Ensure adequate treatment of the underlying sleep disorder while minimizing discomfort symptoms 2
- Objective data: Monitor residual AHI, device usage, and symptom improvement 2
- Close follow-up: After initiation of BPAP by appropriately trained healthcare providers to establish effective utilization patterns and remediate side effects 1
By following these guidelines for switching to BPAP or BPAP ST during PAP desensitization, clinicians can improve patient comfort, adherence, and ultimately treatment outcomes for patients with sleep-disordered breathing.