Pacemaker ECG with Magnet: Recommendations and Procedures
For pacemaker assessment, applying a magnet over the device generator will force asynchronous pacing at a manufacturer-specific rate, which is essential for evaluating capture and battery status, but this should only be done after confirming the specific device's magnet response. 1
Understanding Magnet Effects on Pacemakers
Basic Principles
- Magnet application causes most bradycardia pacemakers to switch to asynchronous pacing (VOO or DOO mode) at a predetermined rate
- The specific magnet-induced pacing rate varies by manufacturer and battery status:
- Normal battery: 85-100 bpm (typical range)
- Decreasing rates indicate battery depletion 1
Important Considerations
- Some pacemaker models have programmable magnet responses or may have this feature disabled 1
- Some devices only force asynchronous pacing for 10 beats before reverting to programmed settings 1
- Always confirm the specific device's magnet response before relying on it for clinical decisions 1, 2
Proper Technique for Magnet Application
- Identify the device type and manufacturer before applying a magnet 1
- Position the magnet directly over the pulse generator (not over the leads)
- Secure the magnet with adhesive tape if continuous asynchronous pacing is needed 1
- Monitor the ECG continuously during magnet application to:
- Confirm effective capture in appropriate chamber(s)
- Assess magnet-induced pacing rate
- Compare current magnet rate with previous values to detect changes 1
Clinical Goals of Magnet ECG Assessment
Without Magnet (Baseline Assessment)
- Determine if patient displays intrinsic rhythm or is being paced
- Characterize underlying atrial mechanism (sinus rhythm, atrial fibrillation, etc.)
- Verify appropriate sensing in relevant chambers 1
During Magnet Application
- Verify effective capture in appropriate chamber(s)
- Assess magnet rate and compare to previous values
- Evaluate pulse width if it's an elective replacement indicator
- Assess threshold margin test results if available and programmed "on"
- Note any changes in AV interval during magnet application (for dual-chamber devices) 1
Special Considerations
Implantable Cardioverter Defibrillators (ICDs)
- Magnet application to an ICD will NOT force asynchronous pacing
- Magnet will only disable the antitachycardia (shocking) function 1
- Some ICD models temporarily suspend shocking function while magnet is in place
- Other models will turn the therapy off completely and require reprogramming 1
- For pacemaker-dependent ICD patients, the device must be programmed to an asynchronous mode (VOO or DOO) 1
Cardiac Resynchronization Therapy (CRT) Devices
- Consider consulting a CIED team for CRT devices as they have more complex programming 1
Leadless Pacemakers
- A magnet will NOT force asynchronous pacing in leadless pacemakers 1
Subcutaneous ICDs
- A properly positioned magnet will emit an R-wave synchronous beep
- If tone is not audible, reprogramming is necessary 1
Common Pitfalls and Caveats
- Failure to identify device type: Always determine the type of CIED, manufacturer, and model before applying a magnet 2
- Inadequate ECG monitoring: Continuous ECG monitoring is essential during magnet application 1
- Relying solely on magnet response: A full device interrogation may be needed if abnormalities are detected 3
- Electromagnetic interference: Can affect ECG interpretation during magnet testing 1
- Obesity or patient positioning: May prevent effective magnet placement in some patients 1
- Misinterpretation of magnet response: Different devices have different responses to magnet application 2
Documentation Requirements
- Record ECG strips showing:
- Baseline rhythm before magnet application
- Response during magnet application
- Return to baseline after magnet removal
- Compare current magnet rate with previous values
- Document any abnormalities in capture or sensing 1
For most routine post-implant assessments, a properly performed magnet check with ECG monitoring may be sufficient without requiring full programmer interrogation 3, but any abnormalities detected during magnet testing should prompt a complete device interrogation.