How to Check Plateau Pressure on the Ventilator Servo I
To check plateau pressure on the Servo I ventilator, use the end-expiratory airway occlusion technique by pressing the inspiratory hold button at the end of inspiration for 0.5-2 seconds and reading the plateau pressure from the ventilator display. 1, 2
Understanding Plateau Pressure
- Plateau pressure reflects alveolar pressure at end-inspiration and is primarily determined by lung/chest wall compliance 1
- Maintaining plateau pressures <30 cmH2O is recommended to prevent ventilator-induced lung injury in ARDS patients 3, 1
- For patients without ARDS, aim for plateau pressures below 25 cmH2O 3
Step-by-Step Procedure for Measuring Plateau Pressure
Prepare the patient:
Perform the measurement:
Interpret the results:
Clinical Applications
- Use plateau pressure to:
Common Pitfalls and Solutions
- Patient-ventilator asynchrony: Ensure the patient is not actively breathing during measurement as this can lead to inaccurate readings 4
- Insufficient hold time: Hold for at least 2 seconds to allow for pressure equilibration throughout the lungs 2
- Circuit leaks: Check for leaks in the ventilator circuit that could lead to falsely low plateau pressure readings 1
- Cardiac oscillations: Small fluctuations in the plateau pressure reading due to cardiac contractions are normal and should not affect interpretation 4