Measuring Aortic and Distal Coronary Pressure During PCI for FFR Calculation
Aortic pressure is measured via a fluid-filled guiding catheter connected to a standard pressure transducer, while distal coronary pressure is measured using a specialized 0.014-inch pressure wire with a sensor located 3 cm from the tip, and FFR is calculated as the ratio of mean distal coronary pressure (Pd) to mean aortic pressure (Pa) during maximal hyperemia induced by intravenous adenosine. 1
Equipment Setup and Preparation
Aortic Pressure Measurement System
- Use a standard fluid-filled pressure transducer connected to the guiding catheter for aortic pressure recordings 1
- Purge all air from the system meticulously and zero the catheter tubing system to atmosphere before starting 1
- Flush the guiding catheter with normal saline at least every 10 minutes to maintain optimal pressure waveform 1
- Remove all contrast medium from the catheter during zeroing steps, as contrast can subtly dampen the pressure waveform 1
Distal Coronary Pressure Measurement System
- Use a 0.014-inch pressure wire with a solid-state electronic sensor mounted 3 cm from the radiopaque tip 1
- Zero the pressure wire sensor ex vivo following manufacturer instructions before insertion 1
- The cross-sectional area of the sensor wire is negligible and does not significantly obstruct flow except in the most critical stenoses 1
Standardized 5-Step FFR Measurement Protocol
Step 1: Zero Both Pressure Systems
- Zero the guiding catheter pressure transducer to atmospheric pressure 1
- Zero the pressure wire sensor ex vivo per manufacturer protocol 1
- Ensure optimal aortic pressure waveform without damping or ventricularization 1
Step 2: Equalize Pressures In Vivo
- Advance the pressure wire sensor a few millimeters beyond the guiding catheter tip 1
- For aorto-ostial lesions, perform equalization with the catheter disengaged in the ascending aorta 1
- Equalize the pressures registered by both the guiding catheter and pressure wire at this position 1
- Document that both pressure signals are identical before proceeding 1
Step 3: Position the Pressure Wire Distally
- Advance the sensor to the distal two-thirds of the coronary artery, at least 2-3 cm distal to the index lesion 1, 2
- Document the final wire position angiographically 1
- Verify the wire tip rotates freely and no resistance is felt when torque is applied 1
- Note: Wire sensor location significantly impacts measurements—more distal positioning yields lower pressure readings 3
Step 4: Induce Maximal Hyperemia and Calculate FFR
- Administer intracoronary nitrates (isosorbide dinitrate 2 mg or nitroglycerin 200 μg) at least 2 minutes before FFR measurement to abolish epicardial vasoconstrictor tone 1
- Infuse intravenous adenosine at 140 μg/kg/min through a large-bore cannula for at least 2 minutes to achieve steady-state maximal hyperemia 1, 2
- Continuous IV adenosine infusion is preferred over intracoronary boluses, which can cause artifacts and provide only transient hyperemia 1
- Calculate FFR as the ratio of mean distal coronary pressure (Pd) to mean aortic pressure (Pa) at the nadir of hyperemia 1, 2
- FFR = Pd/Pa during maximal hyperemia 1, 2
Step 5: Verify for Signal Drift
- Pull the pressure wire back to the guiding catheter tip and verify equal pressures at the end of the procedure 1
- If drift is ≤5 mm Hg, account for this difference in the final FFR calculation 1
- If drift is >5 mm Hg, repeat the entire measurement 1
- Store all measurements on external media for quality review 1
Critical Technical Considerations
Avoiding Catheter-Induced Flow Impairment
- Watch for pressure damping or ventricularization of the aortic pressure waveform, which indicates the guiding catheter is obstructing coronary flow 1
- If damping occurs, gently disengage the guiding catheter from the ostium during measurement while continuing adenosine infusion 1
- This is the most common technical pitfall during FFR measurement 1, 2
Pressure Pullback Technique for Serial Lesions
- For diffuse disease or serial stenoses, perform a slow pullback during sustained hyperemia to identify the location of hemodynamically significant lesions 1
- Focal stenoses show abrupt pressure increases, while diffuse disease shows gradual pressure changes 1
- Obtain an angiogram before pullback and keep the image intensifier position fixed 1
Clinical Interpretation
FFR Thresholds
- FFR ≤0.75 indicates hemodynamically significant stenosis causing ischemia (specificity 100%) 1, 2
- FFR ≥0.80 indicates absence of inducible ischemia (sensitivity 90%) 1, 2
- The "gray zone" between 0.75-0.80 requires clinical judgment 2
Advantages of FFR Over Other Indices
- FFR has high reproducibility and is independent of heart rate, blood pressure, contractility, gender, and CAD risk factors 1, 2
- Unlike resting indices (instantaneous wave-free ratio), FFR requires hyperemia but provides superior diagnostic accuracy 4, 5
- FFR-guided PCI reduces adverse events compared to angiography-guided PCI 2
Common Pitfalls to Avoid
- Never use intracoronary adenosine boluses for FFR measurement—they create artifacts and transient hyperemia 1
- Avoid papaverine or sodium nitroprusside—they provide only transient steady state 1
- In tortuous vessels, the wire itself may create pseudostenoses rendering FFR uninterpretable 1, 2
- Remember that sensor location matters—measurements vary significantly with distance from the lesion, especially near cutoff values 3
- Ensure adequate hyperemia duration (minimum 2 minutes IV adenosine) before recording FFR 1