How is good capture determined in a paced rhythm?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: November 11, 2025View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

Determining Good Capture in a Paced Rhythm

Good capture in a paced rhythm is confirmed when each pacing stimulus produces a corresponding mechanical systole, verified by observing a QRS complex following each pacing spike on ECG and confirming peripheral pulse generation through arterial pressure monitoring, pulse oximetry plethysmogram, or manual pulse palpation. 1

Primary Methods for Assessing Capture

ECG-Based Assessment

  • Verify that each pacing artifact is followed by a QRS complex (for ventricular pacing) or P wave (for atrial pacing) with appropriate morphology 1
  • Measure the paced QRS duration - successful ventricular capture typically produces a wide QRS complex, though left bundle branch pacing may produce narrower complexes (QRS <120 ms) 2
  • Optimize ECG lead selection to minimize pacemaker artifact and maximize QRS visualization, as large pacing artifacts can obscure the QRS complex 1

Critical Pitfall with Transcutaneous Pacing

Because transcutaneous pacing produces large artifacts that may obscure or mimic the QRS complex, making ECG assessment unreliable, you must use concomitant non-ECG monitoring methods (arterial pressure waveform, pulse oximetry plethysmogram, or manual pulse palpation) to confirm mechanical capture 1

Mechanical Confirmation Methods

When ECG assessment is difficult or unreliable:

  • Monitor peripheral pulse through arterial line waveform, pulse oximeter plethysmogram, or manual palpation to confirm each pacing stimulus produces a mechanical systole 1
  • Verify hemodynamic response - adequate blood pressure and perfusion confirm effective mechanical capture 1

Device-Specific Verification

For Permanent Pacemakers

  • Interrogate the device to assess capture thresholds, typically measured in volts at 0.4-0.5 ms pulse width 3, 4
  • Evaluate sensing parameters - appropriate sensing prevents oversensing that could inhibit pacing output 1
  • Check diagnostic counters for loss of capture events or backup pulse delivery, which indicate intermittent capture failure 5

For Biventricular/CRT Devices

  • Confirm biventricular capture by analyzing the paced QRS morphology - look for positive R-wave in V1, as absence may indicate LV lead displacement or loss of LV capture 1
  • Verify 100% biventricular pacing through device diagnostics, as even small percentages of non-captured beats reduce CRT efficacy 1
  • Compare ECG with post-implant template - significant morphology changes suggest loss of capture or lead displacement 1

Threshold Testing for Capture Verification

  • Perform threshold testing by gradually decreasing pacing output until loss of capture occurs, then program output with adequate safety margin 1, 3
  • Maintain at least 100% voltage safety margin above threshold (e.g., if threshold is 1.0V, program output to ≥2.0V) to prevent intermittent capture loss between follow-ups 5
  • Use incremental approach from below threshold during automatic measurements, as beat-to-beat threshold variations can cause underestimation if only 1-2 beats are tested 4

Common Causes of Capture Failure

When capture is lost or intermittent:

  • Lead dislodgement - most common cause, occurring in 1-2% of pacemaker cases and up to 5.7% of CRT cases, with 50% occurring within first 24 hours 1
  • Sudden threshold increase - may indicate lead fracture, insulation breaks, or cardiac perforation 3
  • Oversensing - large P or T waves, muscle artifact, or electrical interference can inhibit pacing output 1
  • Battery depletion or loose connections between lead and generator 1

Monitoring Recommendations

  • Continuous ECG monitoring for 12-24 hours post-implantation to detect early lead dislodgement 3
  • All patients with temporary transvenous pacing require continuous monitoring until device removal or replacement with permanent system 1
  • Check thresholds twice in first 6 months, then annually for single-chamber or every 6 months for dual-chamber devices 3

References

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Research

The electrocardiogram characteristics and pacing parameters of permanent left bundle branch pacing: a systematic review and meta-analysis.

Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing, 2022

Guideline

Pacemaker Threshold Pulse Monitoring Guidelines

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.