Management of Tandem X-2 Insulin Pump and Dexcom CGM During Colonoscopy While NPO
For a patient with diabetes using a Tandem X-2 insulin pump with Dexcom CGM undergoing colonoscopy, continue the insulin pump with basal rate adjustments (reduce to 80% of usual basal or give 60-80% of usual basal dose) and maintain the Dexcom CGM in place throughout the procedure, with hourly point-of-care glucose monitoring as backup. 1
Pre-Procedure Preparation (Days Before)
Perform a basal rate assessment to confirm the pump settings are optimized for fasting conditions 1:
- The patient should fast overnight (after 5-6 PM meal) and monitor glucose every 2-3 hours
- Blood glucose should not vary by more than 1.7 mmol/L (30 mg/dL) during fasting
- If significant deviations occur, contact the diabetes team to adjust basal rates BEFORE scheduling the procedure 1
- This assessment should ideally cover the time period when the colonoscopy is scheduled 1
Coordinate with the patient's diabetes team early to establish a written management plan with signed patient consent to continue pump therapy perioperatively 1
Verify adequate pump supplies: Ensure the patient brings backup infusion sets, insulin cartridges, batteries, and all necessary equipment to cover the entire hospital stay 1
Day Before Colonoscopy
Insulin pump management during bowel preparation:
- Continue the insulin pump with usual basal rates during the bowel prep phase 1
- The patient can manage their own pump settings if alert and capable 1
- Reduce or hold bolus insulin as oral intake decreases 2, 3
- Monitor for hypoglycemia risk during the fasting period, as bowel preparation increases dehydration risk 2, 3
Dexcom CGM considerations:
- Keep the Dexcom sensor in place throughout preparation 1
- Be aware that dehydration and vasoconstriction during bowel prep may affect CGM accuracy 1
- Confirm point-of-care glucose readings if CGM values seem inconsistent 1
Day of Colonoscopy
Insulin pump settings during NPO period:
- Continue basal insulin at 80% of the usual rate as a temporary basal setting once fasting begins 1
- Alternatively, some protocols recommend 60-80% of usual long-acting insulin equivalent if converting to injections 1
- Do NOT completely discontinue insulin - patients using only rapid-acting insulin in pumps become insulin-deficient within 1-4 hours of cessation, risking hyperglycemia and ketosis 1
- Hold all bolus insulin doses while NPO 1
Glucose monitoring protocol:
- Monitor blood glucose with point-of-care testing every 4-6 hours minimum while NPO 1
- Continue Dexcom CGM for trend monitoring, but do not rely solely on CGM values - always confirm with fingerstick or point-of-care testing 1, 4
- Target glucose range: 80-180 mg/dL (4.4-10.0 mmol/L) perioperatively 1
Infusion site considerations:
- Position the infusion site away from the surgical field (typically abdomen, thighs, or buttocks) 1
- If diathermy will be used, switch to a non-metal infusion set in advance 1
- Ensure the site is accessible to the anesthesia team throughout the procedure 1
During the Colonoscopy Procedure
Pump continuation criteria - the pump can remain on IF 1:
- The patient signed consent for continued pump use
- Hospital policies support inpatient pump use
- The diabetes team is available for consultation
- The procedure is relatively short (colonoscopy typically qualifies)
- The patient will regain consciousness shortly after the procedure
Contraindications requiring pump discontinuation 1:
- Prolonged impaired consciousness beyond brief anesthesia
- Patient inability to self-manage upon awakening
- Lack of trained healthcare providers to manage the pump
- Hyperglycemic crisis (DKA)
- Equipment malfunction or lack of supplies
If pump must be discontinued, convert to basal-bolus subcutaneous insulin using the 24-hour total basal dose from pump settings to calculate long-acting insulin dose 1
CGM management during procedure:
- The Dexcom CGM can remain in place during colonoscopy 1, 4
- Remove CGM only if MRI or specific electromagnetic procedures are planned 1
- Continue hourly point-of-care glucose monitoring during the procedure 1, 4
Post-Procedure Recovery
Resuming normal pump function:
- Return to 100% basal rate once the patient is awake and alert 1
- Resume bolus insulin with the first meal or snack 1
- The patient should resume self-management of pump settings as soon as they are capable 1
Glucose monitoring:
- Continue point-of-care testing every 4-6 hours until oral intake is well-established 1
- The Dexcom CGM provides continuous monitoring but should not replace scheduled fingerstick checks initially 1, 4
- Address hypoglycemia (<70 mg/dL) immediately with 15g fast-acting carbohydrate once oral intake permitted 1
- For hyperglycemia >300 mg/dL (16.7 mmol/L), check for ketones and administer correction bolus via pump 1
Critical Safety Considerations
Common pitfalls to avoid:
- Never completely discontinue insulin in pump users - they have no long-acting insulin on board 1, 5
- Do not rely solely on CGM values during acute physiological changes (dehydration, hypotension) 1
- Avoid removing CGM unnecessarily, as this disrupts the Tandem Control-IQ closed-loop system and may cause prolonged interruption of automated insulin delivery 1, 4
- Do not assume hospital staff can manage the pump - ensure the patient or diabetes team maintains control 1
Emergency protocols:
- If the pump malfunctions, immediately convert to subcutaneous basal-bolus insulin 1
- If severe hyperglycemia (>300 mg/dL) with ketones develops, initiate DKA protocol and discontinue pump 1
- For hypoglycemia during sedation, treat with IV dextrose and reduce basal rate by 20-50% 1
Documentation requirements: