Do implantable cardiac defibrillators (ICDs) need to be turned off during surgery?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: November 24, 2025View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

ICD Management During Surgery

Yes, implantable cardioverter defibrillators must have their tachyarrhythmia treatment algorithms programmed off before surgery to prevent inappropriate shocks from electromagnetic interference that the device could misinterpret as ventricular tachycardia or fibrillation. 1

Preoperative Device Management

The preferred method is formal reprogramming using an ICD programmer to disable the shock function before surgery, rather than relying on magnet placement alone. 1, 2 This approach provides more reliable control because:

  • Magnet effects vary significantly by device manufacturer and model 1
  • Some ICD models only temporarily suspend shock therapy while the magnet is in place, and therapy reactivates immediately upon magnet removal (whether intentional or accidental) 1
  • Other models may permanently disable therapy with magnet application, requiring reprogramming to restore function 1

Device evaluation should occur within 3-6 months before significant surgical procedures, particularly for major abdominal or thoracic surgery involving substantial electrocautery. 1, 2

Critical Consideration for Pacemaker-Dependent Patients

If the patient is pacemaker-dependent, the pacing function of the ICD must be reprogrammed to an asynchronous mode (VOO or DOO) during surgery to prevent electromagnetic interference from inhibiting necessary pacing. 1, 2 This is a separate issue from disabling shock therapy and requires specific attention, as some ICD patients rely on the device for both defibrillation and pacing support.

Rationale: Prevention of Inappropriate Shocks

The primary reason for disabling ICD therapy is that electrocautery and other surgical equipment generate spurious electrical signals that ICDs can misinterpret as life-threatening ventricular arrhythmias, triggering inappropriate shocks. 1 These inappropriate shocks:

  • Cause significant patient harm and distress during surgery
  • Can be arrhythmogenic themselves 3
  • Severely impair quality of life 3, 4
  • Occur with concerning frequency even outside the surgical setting, making prevention during high-risk EMI exposure essential 5, 3, 6

Intraoperative Precautions

Even with the ICD disabled, implement these protective measures 2:

  • Use bipolar electrocautery instead of monopolar when possible (lower EMI risk)
  • Apply short, intermittent bursts of electrocautery at the lowest effective energy
  • Maximize distance between the electrocautery site and the ICD generator/leads
  • Maintain continuous ECG and pulse monitoring (recognizing that electrocautery may interfere with ECG display)

Emergency Cardioversion During Surgery

If emergent cardioversion becomes necessary during surgery, place defibrillation paddles as far from the ICD as possible, using anterior-posterior paddle positioning perpendicular to the device leads. 1 This orientation minimizes potential damage to the ICD system.

Postoperative Device Restoration

After surgery, the ICD's antitachycardia function must be restored to active status before hospital discharge. 1, 2 This critical step requires:

  • Formal device interrogation using a programmer
  • Verification through a printed interrogation report confirming antitachycardia therapy is active 1
  • Assessment of overall device function, particularly if the device was reprogrammed preoperatively 2

Common Pitfalls to Avoid

Never assume magnet placement alone provides adequate ICD management during surgery. 1 The variable and unpredictable effects of magnets across different device models make formal reprogramming the only reliable approach.

Do not forget to reactivate the ICD after surgery. 1, 2 Patients have been discharged without functional defibrillation protection, creating life-threatening vulnerability. A documented interrogation printout provides verification and medicolegal protection.

Remember that disabling shock therapy does not address pacemaker dependency. 1, 2 These are separate functions requiring independent management decisions based on the patient's underlying rhythm and pacing requirements.

References

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Guideline

Perioperative Management of Patients with Cardiac Implantable Electronic Devices

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Research

Tools and strategies for the reduction of inappropriate implantable cardioverter defibrillator shocks.

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2008

Research

[Management of inappropriate shocks/T-wave-oversensing in S-ICD®-patients].

Herzschrittmachertherapie & Elektrophysiologie, 2018

Research

Inappropriate implantable cardioverter-defibrillator discharges unrelated to supraventricular tachyarrhythmias.

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2006

Related Questions

What is the indicated management for a patient with an automatic implantable cardioverter-defibrillator (AICD) presenting with weakness, palpitations, tachycardia, and hypotension?
Does an implanted defibrillator (implantable cardioverter-defibrillator, ICD) cause scarring and further fibrosis of the myocardium every time it discharges?
Does the implanted defibrillator's (Implantable Cardioverter-Defibrillator, ICD) charging process affect the myocardium?
What are the potential side effects and complications of an Implantable Cardioverter-Defibrillator (ICD)?
Is it safe for individuals with subcutaneous implantable cardioverter-defibrillators (ICDs) to use hot tubs?
What is the immediate treatment for a patient with a lacunar infarct?
What hepatitis B (HBV) vaccine is recommended for an adult undergoing chemotherapy?
What is the appropriate workup for elevated liver enzymes (elevated Alanine Transaminase (ALT), Aspartate Transaminase (AST), alkaline phosphatase, and total and direct bilirubin)?
Does a patient with low iron saturation (hypoferritinemia) of 10% require iron supplementation if iron and ferritin levels are normal?
What is the most appropriate next step for a 16-year-old boy who presented to the emergency department (ED) after a syncopal episode with a history of palpitations, chest pain, and a family history of sudden cardiac death?
What is the best course of action for an elderly patient on losartan (angiotensin II receptor antagonist) 25mg daily with persistent hypertension?

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.