Is the Medtronic (Medical Technology Company) X2DR01 Astra (Implantable Medical Device) MRI (Magnetic Resonance Imaging) compatible?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: December 7, 2025View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

MRI Compatibility of Medtronic X2DR01 Astra Device

The Medtronic X2DR01 Astra labeled as "MRI conditional" is compatible with MRI scanning, but only when specific manufacturer-defined conditions are strictly followed—this is not the same as being universally "MRI safe."

Understanding MRI Conditional Status

MRI conditional devices can be safely scanned only when all manufacturer specifications are met, including field strength limitations, specific anatomical exclusion zones, scanning duration limits, and required device programming 1, 2.

Critical Distinctions Between Device Categories

  • MRI conditional devices require adherence to specific scanning parameters defined by the manufacturer, including magnetic field strength (typically 1.5T or 3T), spatial gradient limits, and specific absorption rate (SAR) restrictions 1, 2
  • Not all MRI conditional devices are created equal—some manufacturers limit scanning to 30 minutes, require exclusion zones in the thorax, or only approve scanning in afebrile patients 2
  • Device approval varies by field strength—not all companies have devices approved for both 1.5T and 3T scanners 2

Pre-Scan Requirements for MRI Conditional Devices

Device Verification and Component Assessment

  • Every component of the implanted system must be identified and verified as MRI conditional—the most restrictive component dictates the entire scanning approach 3
  • Confirm that all leads, generators, and accessories are MRI conditional, as a single non-conditional component makes the entire system unsafe 3, 4
  • Verify the specific model numbers against manufacturer MRI conditional labeling, as the CapSureFix Novus 5076 lead has been studied with MRI conditional pacemakers 5

Mandatory Device Programming

  • Specialized MRI mode programming is essential and must be individualized based on the patient's underlying rhythm 6
  • Program the device to asynchronous pacing mode (DOO/VOO) to prevent inappropriate inhibition during MRI scanning 1, 2
  • Three documented cases showed competing paced beats during MRI when programming was not optimally selected, including situations where intrinsic heart rate increased or atrial fibrillation occurred during scanning 6

Technical Modifications in MRI Conditional Devices

The X2DR01 Astra's MRI conditional status depends on specific engineering modifications 2:

  • Reduction in ferromagnetic components to minimize magnetic attraction and susceptibility artifacts
  • Replacement of reed switch with Hall sensor to prevent unpredictable switch behavior during scanning
  • Modified lead coil design to minimize heating and electrical current induction
  • Filter circuitry protecting the internal power supply from electromagnetic interference
  • Dedicated programming algorithms to prevent inappropriate pacemaker inhibition and competing rhythms

Scanning Protocol Requirements

Field Strength and Technical Parameters

  • Verify the approved magnetic field strength (1.5T vs 3T) for this specific device model 2, 4
  • Confirm spatial gradient and SAR limits do not exceed manufacturer specifications 1, 2
  • Some devices restrict scanning duration to 30 minutes—verify time limitations for the X2DR01 Astra 2

Anatomical Restrictions

  • Check for thoracic exclusion zones—not all MRI conditional devices allow unlimited scanning without anatomical restrictions 2
  • Certain manufacturers limit the imaging field of view or require specific patient positioning 2, 4

Patient-Specific Contraindications

  • Scanning may be restricted to afebrile patients only depending on manufacturer specifications 2
  • Verify patient temperature before proceeding with MRI examination 2

Post-Scan Device Management

  • Reprogram the device to normal settings immediately after MRI completion 4, 6
  • Perform threshold testing and sensing checks to verify normal device function 5, 4
  • Document any changes in pacing capture thresholds (acceptable change ≤0.5V) and sensing amplitudes 5

Critical Safety Warnings

MRI conditional does not mean universally safe—serious device-related MRI safety incidents have been reported when conditions are not strictly followed 3, 4.

Common Pitfalls to Avoid

  • Never assume all components are MRI conditional without verification—a single non-conditional lead makes the entire system unsafe 3
  • Do not proceed with scanning if any manufacturer condition cannot be met 2, 4
  • Avoid generic MRI mode programming—competing rhythms occurred in documented cases when programming was not carefully selected for the individual patient 6
  • Do not scan patients with conventional (non-MRI conditional) devices unless following rigorous research protocols with appropriate safety measures 1, 4

Practical Implementation

Before authorizing MRI for a patient with the Medtronic X2DR01 Astra:

  1. Obtain complete device information including all component model numbers 3
  2. Verify MRI conditional status of every component against manufacturer documentation 3, 4
  3. Confirm the MRI scanner specifications match approved parameters (field strength, SAR, gradient) 2, 4
  4. Program device to appropriate MRI mode based on patient's underlying rhythm 6
  5. Verify no anatomical or patient-specific contraindications exist 2
  6. Ensure post-scan device interrogation and reprogramming capability is available 5, 4

The widespread adoption of MRI conditional devices depends on real-world factors including cost, clear clinical indications, and permanent education of healthcare professionals 2.

References

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Research

Magnetic resonance imaging-conditional devices: Luxury or real clinical need?

Hellenic journal of cardiology : HJC = Hellenike kardiologike epitheorese, 2017

Research

MR Safety: Active Implanted Electronic Devices.

Magnetic resonance imaging clinics of North America, 2020

Research

MRI and cardiac implantable electronic devices; current status and required safety conditions.

Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation, 2014

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.