Reasons for False Glucometer Readings
Glucometer inaccuracies stem from three primary categories: operator errors (the most significant source), physiological/patient factors, and environmental/technical factors, with healthcare providers needing to verify discordant readings through laboratory testing and select appropriate meter technology based on patient-specific conditions. 1, 2
Operator and User-Related Errors (Most Common Source)
Operator errors exceed instrument errors as the primary cause of inaccurate readings 1, 2:
- Failure to calibrate meters regularly is one of the most common preventable errors 1, 2
- Improper hand washing leading to glucose contamination on fingertips (from food, lotions, or other glucose-containing products) 2, 3
- Dirty or poorly maintained meters affecting sensor function 2
- Improper test strip storage exposing strips to humidity or excessive temperature, which can falsely elevate results 1
- Using expired test strips or strips purchased from unlicensed distributors 1
- Improper use of control solutions for quality checks 2
- Inadequate patient training on proper technique 1
Physiological and Patient-Related Factors
Hematocrit Abnormalities
Low hematocrit (<35%) causes falsely elevated glucose readings on most point-of-care meters, particularly glucose oxidase-based systems 1, 4:
- Anemia from any cause (iron deficiency, chronic kidney disease, acute blood loss) 4
- Hemodilution from fluid resuscitation or volume expansion 4
- A patient with true glucose of 80 mg/dL and hematocrit of 0.25 may show readings up to 18 mg/dL higher 4
Oxygen Tension Abnormalities
Glucose oxidase monitors are highly sensitive to oxygen levels and should only be used in patients with normal oxygen saturation 1:
- Higher oxygen tension (arterial blood, oxygen therapy) causes falsely LOW glucose readings 1
- Lower oxygen tension (high altitude, hypoxia, venous blood) causes falsely HIGH glucose readings 1
- Glucose dehydrogenase monitors are not oxygen-sensitive and should be selected for patients with abnormal oxygenation 1
Interfering Medications and Substances
For glucose oxidase monitors specifically 1:
For glucose dehydrogenase monitors 1:
- Icodextrin (used in peritoneal dialysis) 1
Additional interfering factors include dopamine, mannitol, and high triglyceride concentrations 1, 5
Environmental and Technical Factors
Temperature Effects
All glucose monitors have acceptable temperature ranges, and readings outside these ranges produce errors 1:
- Cold temperatures lower the glucose result, causing euglycemic levels to read as hypoglycemic and hyperglycemic levels to appear better than actual 6
- Hot temperatures increase the glucose result, causing hypoglycemic levels to falsely read as euglycemic 6
- Most meters display error messages when temperature is unacceptable, but some provide readings with warnings that values may be incorrect 1
Altitude and Humidity
- High altitude affects oxygen-dependent glucose oxidase meters 1
- High humidity (>80%) can affect strip integrity and meter performance 3, 6
Alternate Site Testing
Forearm and thigh blood glucose testing is unreliable during periods of rapid glucose change 1:
- After meals, forearm blood glucose rises more slowly and less high than fingertip blood 1
- After exercise, thigh and forearm glucose levels fall lower than fingertip glucose 1
- Fingertip testing is mandatory when glucose is changing rapidly (post-meal, post-exercise, during insulin treatment, suspected hypoglycemia) 1
Critical Care and Special Population Considerations
In critically ill patients, glucometers have particularly high error rates 2:
- Avoid glucometers in patients with shock, on vasopressors, or with peripheral edema 2
- Use arterial or venous whole blood sent to the laboratory with hexokinase-based methods 2
- Laboratory-based values should be used if there is any doubt about meter accuracy 1
In neonates 2:
- High hematocrit common in newborns significantly affects accuracy 2
- Use meters specifically intended for neonatal use 2
- Analytical bias at low concentrations causes frequent false alarms or missed hypoglycemia 2
Meter-Specific Accuracy Limitations
Many widely-used meters fail to meet FDA/ISO accuracy standards in real-world practice 1, 2:
- FDA requires 95% of readings within 15% of laboratory results for home-use meters 2
- A recent study found only 6 of the top 18 glucose meters met defined accuracy standards 1
- Meters are less reliable in the lower ranges of glycemia (hypoglycemia) 1
- Meters may overestimate true glucose values in the high glycemic range 1
- Variation exists even between two meters of the same brand from a single manufacturer 1
Clinical Approach to Suspected Inaccurate Readings
When a meter reading seems discordant with clinical reality, it must be retested or verified with laboratory testing 1:
- Suspect inaccuracy when home glucose testing is inconsistent with A1C levels 1
- Patients can use control solutions to check meter accuracy if equipment malfunction is suspected 1
- Compare SMBG with concurrent laboratory glucose analysis at regular intervals 1
- Reevaluate patient technique at each clinic visit 1
Practical Recommendations for Meter Selection
Choose meter technology based on patient-specific factors 1, 2, 4:
- Select glucose dehydrogenase meters for patients with abnormal oxygen saturation, chronic lung disease, or those on oxygen therapy 1, 2
- Avoid glucose oxidase-based meters in patients with hematocrit <35% or abnormal oxygenation 4
- Use only FDA-approved meters with unexpired strips purchased from licensed distributors 1
- Use professional-grade meters in hospital and acute care settings that meet higher accuracy standards (95% within 12 mg/dL for glucose <100 mg/dL or 12% for glucose ≥100 mg/dL) 2