Converting a Closed System to a Semi-Closed System in an Anesthesia Machine
To convert from a closed to a semi-closed system, simply increase the fresh gas flow (FGF) rate above the patient's metabolic consumption—typically setting FGF to 1-3 L/min—and ensure the adjustable pressure limiting (APL) valve is partially open to allow excess gas to escape through the scavenging system. 1
Understanding the Key Difference
The fundamental distinction between closed and semi-closed systems is the fresh gas flow rate relative to patient uptake:
- Closed system: FGF equals exactly what the patient consumes metabolically (typically 200-300 mL/min O₂), with no excess gas vented 2
- Semi-closed system: FGF exceeds patient uptake, with surplus gas eliminated via the APL valve into the scavenging system 2, 3
Step-by-Step Conversion Process
1. Adjust Fresh Gas Flow Rate
- Increase FGF to 1-3 L/min (or higher) from the minimal closed-circuit flow 3
- At FGF of 1 L/min, you achieve "minimal flow anesthesia," which is effectively semi-closed 2
- At FGF of 3 L/min or greater, you have a standard semi-closed system 3
- Higher FGF rates (5-10 L/min) provide faster control of anesthetic depth but increase costs and waste 4, 3
2. Configure the APL Valve
- Ensure the APL valve can be fully opened to allow excess gas to escape 1
- Verify the valve is functioning properly by squeezing the reservoir bag during manual ventilation 1
- The APL valve must be connected to the scavenging system to prevent operating room pollution 1
3. Verify System Integrity
- Perform a two-bag test to confirm proper circuit function 1:
- Attach a test lung to the patient end
- Set FGF to 5 L/min and ventilate manually
- Check that unidirectional valves are moving (if present)
- Test APL valve function by squeezing both bags
- Verify no volume loss when vaporizers are opened/closed with ventilator running 1
4. Confirm Scavenging System Function
- Verify the scavenging system is switched on and functioning 1
- Ensure tubing is properly attached to the exhaust port of the breathing system 1
- Check for leaks in gas sampling lines, as these are common sources of system leaks 1
Clinical Considerations
Advantages of Semi-Closed vs. Closed Systems
- Easier control of anesthetic depth: Higher FGF allows faster changes in inspired anesthetic concentration 4, 2
- No need for continuous adjustment: Unlike closed systems that require constant manual titration of gas delivery 2
- More forgiving: Less risk of hypoxia or awareness from calculation errors 2
- Standard practice: Closed systems are not routine due to technical complexity 2
Flow Rate Selection Strategy
- 0.5-1 L/min (minimal flow): Maximizes economy and humidity, requires careful monitoring 2, 3
- 1-3 L/min: Balances cost savings with ease of control 3, 5
- 3-5 L/min: Standard semi-closed approach, easier anesthetic control 3
- >5 L/min: Fastest control but highest cost and waste 3
Common Pitfalls to Avoid
- Forgetting to open the APL valve: This can cause dangerous pressure buildup in a closed system when FGF is increased 1
- Inadequate scavenging: Increased FGF without proper scavenging leads to operating room pollution 1
- Ignoring CO₂ absorbent: Semi-closed systems still require functional CO₂ absorption; check absorbent color and supply 1, 6
- Assuming immediate concentration changes: Even with higher FGF, time to target concentration varies by machine type and internal volume 4