Measuring Oxygen Flow Rate with Fully Open Valve
When an oxygen valve is fully open, the oxygen flow rate is measured using a calibrated flow meter or flow analyzer device that directly quantifies the volume of oxygen delivered per minute (L/min), not by the valve position itself. 1
Understanding Flow Measurement vs. Valve Position
The valve position (fully open or partially open) does not directly indicate the actual flow rate being delivered. The flow rate must be measured using proper equipment:
- Flow meters attached to oxygen sources (cylinders, concentrators, or wall outlets) display the set flow rate, but their accuracy varies significantly 2
- Calibrated flow analyzers (such as the Timeter RT-200) provide the most accurate measurement of actual oxygen delivery 3, 2
- The oxygen device and flow rate should be documented on arrival at hospital or when oxygen therapy becomes necessary 1
Critical Measurement Considerations
Flow Meter Accuracy Issues
Flow meters commonly used in hospital settings show poor accuracy despite good precision, particularly at low flow rates (1-3 L/min) where variability is greatest 2:
- At 1 L/min: measured flows are typically lower than the set rate 2
- At 3 L/min: measured flows are close to the set rate 2
- At 5-10 L/min: measured flows are typically higher than the set rate 2
- Large variability exists between different flow meters, even when new 2
Factors Affecting Actual Delivered Flow
The actual oxygen flow reaching the patient depends on multiple system variables beyond the valve setting 1, 4:
- Tubing length: Flow reductions become clinically significant (>20%) with cylinder systems at tubing lengths exceeding 100 feet (30.48 m) at 2 L/min 4
- Oxygen source type: Liquid oxygen systems maintain flow better than cylinders or concentrators with extended tubing 4
- Connection points: Oxygen attached to the common gas outlet delivers more accurately than Y-piece connections 3
Monitoring Oxygen Delivery
Essential Monitoring Parameters
Rather than relying solely on flow meter readings, clinical monitoring should focus on the patient's oxygen saturation (SpO2) as the primary indicator of adequate oxygen delivery 1:
- Check SpO2 within 5 minutes of starting or changing oxygen therapy 1
- Monitor continuously for critically ill patients 1
- Stable patients require four-times-daily monitoring 1
- Target saturation ranges: 94-98% for most patients or 88-92% for those at risk of hypercapnic respiratory failure 1, 5
Verification Steps When Flow Issues Suspected
If oxygen saturation falls below target despite a fully open valve, systematically verify 1:
- Oximeter is correctly placed and functioning
- Oxygen delivery device is appropriate and properly positioned
- Flow rate setting matches clinical requirements
- For cylinders: verify it is an oxygen cylinder (check labeling) and not empty or near-empty 1
- Check all connections for leaks or disconnections 1
Device-Specific Flow Characteristics
Nasal Cannula
Nasal cannula flow rates cannot predict precise FiO2 delivery, as actual inspired oxygen concentration varies with patient breathing patterns 1, 5:
- Each L/min increases FiO2 by approximately 3-4% above room air (21%) 5
- High respiratory rates reduce effective FiO2 5
- Mouth breathing reduces FiO2 by an average of 2.4% 5
- Flows above 4 L/min may cause patient discomfort 1, 5
Venturi Masks
Venturi masks provide the most predictable oxygen delivery with specific flow rates for each concentration 1, 5:
- 24% oxygen: 2 L/min input, total flow 51 L/min 1
- 28% oxygen: 4 L/min input, total flow 44 L/min 1
- Increasing oxygen flow into a Venturi mask does not increase the oxygen concentration delivered 1
- For patients with respiratory rate >30 breaths/min, set flow above the minimum listed on packaging 1
Clinical Pitfalls
- Never assume the flow meter reading equals actual delivered flow—verify with patient response (SpO2) 1, 2
- Do not use simple face masks below 5 L/min due to CO2 rebreathing risk 5
- Extended oxygen tubing significantly reduces flow from cylinder systems; verify adequate delivery at the patient end 4
- Flow meter readings at high settings (10 L/min) show the greatest inaccuracy between new and used equipment 2