Diabetes Diagnosis Using A1C
An A1C ≥6.5% measured by an NGSP-certified laboratory assay is diagnostic for diabetes, but requires confirmation with a repeat test unless the patient has unequivocal hyperglycemia. 1
Diagnostic Criteria
The American Diabetes Association recognizes A1C ≥6.5% as one of four acceptable diagnostic criteria for diabetes: 1
- A1C ≥6.5% (48 mmol/mol) using an NGSP-certified, DCCT-standardized laboratory method
- Fasting plasma glucose ≥126 mg/dL (7.0 mmol/L) after at least 8 hours of fasting
- 2-hour plasma glucose ≥200 mg/dL (11.1 mmol/L) during oral glucose tolerance test
- Random plasma glucose ≥200 mg/dL (11.1 mmol/L) in patients with classic hyperglycemic symptoms
In the absence of unequivocal hyperglycemia, diagnosis requires two abnormal test results from the same sample or in two separate test samples. 1, 2
Confirmation Strategy
When using A1C for diagnosis, follow this algorithmic approach: 2
- Repeat the same test (A1C) for confirmation, as this provides the greatest likelihood of concordance 2
- If two different tests (e.g., A1C and fasting glucose) are both above diagnostic thresholds, the diagnosis is confirmed without further testing 2
- If results are discordant between two different tests, repeat the test that exceeded the diagnostic threshold 2
- For results near diagnostic margins (e.g., A1C 6.4-6.6%), repeat testing in 3-6 months with close clinical follow-up 2
Example: If initial A1C is 7.0% and repeat is 6.8%, diabetes is confirmed even though the second value is lower, as both exceed 6.5%. 2
Critical Testing Requirements
Laboratory Standards
A1C must be performed using an NGSP-certified method standardized to the DCCT assay to avoid misdiagnosis. 1
- Point-of-care A1C testing is not recommended for diagnosis due to lack of standardization 3
- Point-of-care testing may be used for monitoring established diabetes and allows timely treatment adjustments 3
- Clinical laboratories must be CLIA-certified with documented personnel competency and proficiency testing 1
Practical Advantages
A1C offers significant practical benefits over glucose-based testing: 3
- No fasting required - can be drawn at any time of day
- No meal timing restrictions - unaffected by recent food intake or exercise
- Greater preanalytic stability - less temporal variability than glucose measurements 4
- Reflects average glycemia over approximately 2-3 months 1
When A1C Should NOT Be Used for Diagnosis
In conditions with altered red blood cell turnover or hemoglobinopathies, use only plasma glucose criteria for diagnosis. 1
Absolute Contraindications to A1C Diagnosis
Use plasma glucose testing exclusively in: 1
- Hemoglobinopathies including sickle cell disease (not just trait)
- Pregnancy (second and third trimesters and postpartum period)
- Glucose-6-phosphate dehydrogenase deficiency
- HIV infection (A1C may underestimate glycemia) 1
- Hemodialysis
- Recent blood loss or transfusion
- Erythropoietin therapy
Race/Ethnicity Considerations
African Americans with sickle cell trait may have A1C values approximately 0.3% lower than their true glycemic burden. 1
- The G6PD G202A variant (present in 11% of African Americans) can lower A1C by 0.7-0.8% 1
- African Americans may have higher A1C levels than non-Hispanic Whites at similar glucose levels 1
- When marked discordance exists between A1C and plasma glucose, use an assay without hemoglobin variant interference or rely on glucose criteria 1
Diagnostic Performance Considerations
A1C ≥6.5% identifies fewer individuals with diabetes compared to glucose-based criteria, particularly missing those with isolated postprandial hyperglycemia. 5, 4
The 2-hour post-glucose load diagnoses more people with diabetes than either fasting glucose or A1C 1. When A1C and glucose tests are discordant, fasting and 2-hour plasma glucose are more accurate 1.
Test Performance Data
Against repeated fasting glucose measurements, A1C ≥6.5% demonstrates: 6
- Sensitivity: 67% (misses approximately one-third of diabetes cases)
- Specificity: 97% (very few false positives)
- Area under the curve: 0.936
Individuals with both elevated fasting glucose ≥126 mg/dL AND A1C ≥6.5% have an 88% 10-year risk of diagnosed diabetes, compared to 55% with elevated fasting glucose alone. 6
Common Pitfalls to Avoid
- Never use point-of-care A1C for initial diagnosis - only NGSP-certified laboratory assays are acceptable 3
- Don't ignore discordance - when A1C and glucose results conflict, investigate for conditions affecting red blood cell turnover 1, 2
- Ensure proper sample handling - plasma glucose samples must be spun and separated immediately after collection to prevent falsely low results 2
- Don't assume A1C is always reliable - in populations with high hemoglobinopathy prevalence (e.g., Tanzania with HIV), A1C may systematically underestimate glycemia 1
- Avoid mixing diagnostic methods inappropriately - if using A1C for diagnosis, confirm with repeat A1C rather than switching to glucose testing unless there's clinical reason for discordance 7
Monitoring After Diagnosis
Once diabetes is diagnosed, A1C frequency depends on glycemic control: 1