Proper ECG Electrode Placement
For diagnostic 12-lead ECGs, place limb electrodes on the arms and legs distal to the shoulders and hips (traditionally at wrists and ankles), and position precordial electrodes V1-V6 using precise anatomical landmarks with the patient supine. 1
Limb Electrode Placement
Standard Diagnostic ECG
- Place the four limb electrodes on the wrists and ankles with the patient in the supine position 1
- The American Heart Association recommends placement on the arms and legs distal to the shoulders and hips, though not necessarily at the exact wrists and ankles 1
- Electrode placement along the limbs significantly affects ECG voltages and durations, particularly in limb leads 1
Exercise/Treadmill Testing (Modified Placement)
- Use the Mason-Likar lead position to reduce motion artifact during exercise testing 2
- Place arm electrodes in the infraclavicular fossae medial to the deltoid insertions or over the outer clavicles 2
- Position the left leg electrode midway between the costal margin and iliac crest in the left anterior axillary line 2
- ECGs recorded with torso placement cannot be considered equivalent to standard ECGs and must not be used interchangeably for serial comparison 2, 3
Precordial Electrode Placement (V1-V6)
Precise Anatomical Positions
- V1: Fourth intercostal space at the right sternal border 1
- V2: Fourth intercostal space at the left sternal border 1
- V3: Midway between V2 and V4 1
- V4: Fifth intercostal space in the midclavicular line 1
- V5: In the horizontal plane at the anterior axillary line (same horizontal level as V4) 1
- V6: In the horizontal plane at the midaxillary line (same horizontal level as V4) 1
Critical Technical Points
- Position precordial electrodes with reference to underlying bony landmarks to avoid an erroneously vertical orientation 1, 4
- V5 and V6 should follow the horizontal plane of V4, not the fifth intercostal space, as the intercostal space course is variable 1
- Avoid using the "anterior axillary line" as a marker for V5, as its definition is vague 1
Skin Preparation
- Shave the electrode application areas, then rub with alcohol-saturated gauze 1
- After the skin dries, mark with a felt-tipped pen and rub with fine sandpaper or rough material 1
- Skin preparation by cleaning and gentle abrasion reduces noise and improves ECG quality 1
- These procedures should reduce skin resistance to 5000 Ω or less 1
Common Pitfalls and Their Consequences
Superior Misplacement of V1 and V2
- Placing V1 and V2 in the second or third intercostal space reduces initial R-wave amplitude by approximately 0.1 mV per interspace 1, 4
- This causes poor R-wave progression or erroneous signs of anterior infarction 1, 4
- Superior displacement often produces rSr' complexes with T-wave inversion resembling lead aVR 1
- Only 16% of cardiologists correctly identified V1 position in one study, compared to 90% of cardiac technicians 5
Inferior Misplacement of V5 and V6
- Placing V5 and V6 in the sixth intercostal space or lower alters amplitudes used for ventricular hypertrophy diagnosis 1, 4
- This can cause false-negative results for left ventricular hypertrophy 4
Torso Placement of Limb Electrodes
- Moving limb electrodes from extremities to torso causes rightward axis shifts and voltage changes 6
- Over half (13 of 25) of ECGs showing inferior infarct criteria with standard placement had those criteria erased with Mason-Likar positioning 6
- R wave amplitude increases >3 mm in inferior leads can cause disappearance of inferior infarcts 7
- Decreases >3 mm in leads I and aVL may cause false lateral infarcts 7
Special Populations
- In patients with obstructive pulmonary disease and low diaphragm position, V3 and V4 may record negative deflections that falsely simulate anterior infarction 1, 4
- In women with large breasts, place electrodes beneath the breast to reduce amplitude attenuation and favor reproducibility 1
Documentation Requirements
- Any ECG using modified electrode placement must be clearly labeled as such to prevent misinterpretation when compared to previous standard ECGs 2
- Record any modification of lead placement on the tracing 1