How do I calculate physiological dead space using the Bohr equation?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: February 28, 2026View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

How to Calculate Physiological Dead Space

Physiological dead space is calculated using the Bohr equation: VD/VT = (PaCO2 - PECO2)/PaCO2, where PaCO2 is arterial CO2 and PECO2 is mixed expired CO2. 1, 2

Required Measurements

To calculate dead space, you need two essential values:

  • Arterial PCO2 (PaCO2): Obtained from arterial blood gas analysis 2
  • Mixed expired PCO2 (PECO2): Measured by collecting expired gas in a large gas-impermeable bag over time and analyzing the PCO2 of the collected gas, or by sampling from a mixing chamber in modern exercise systems 1, 2

For breath-by-breath systems, PECO2 can be calculated from the VCO2/VE ratio, where both gas volumes are expressed in liters (STPD) 1, 2

The Bohr Equation

The standard Bohr equation is:

VD/VT = (PaCO2 - PECO2) / PaCO2 1, 2

This yields a dimensionless fraction representing the proportion of each tidal volume that does not participate in gas exchange 1

Modified Equation for Mechanical Ventilation

When patients are connected to ventilator equipment with additional dead space from tubing and valves, use the modified equation:

VD/VT = [(PaCO2 - PECO2) / PaCO2] + (VDvalve/VT) 1, 2

Where VDvalve represents the dead space of the mouthpiece, valve, and ventilator circuit 1, 2

Important Technical Considerations

Do Not Use End-Tidal CO2 as a Substitute

End-tidal PCO2 (PETCO2) should NOT be used as an index of PaCO2, especially in patients with lung disease. 1 This is misleading because PETCO2 can actually exceed PaCO2 during exercise in normal individuals, and there are no valid procedures to adequately estimate PaCO2 noninvasively in patients with lung disease 1

Equipment Dead Space Limits

Total apparatus dead space should not exceed 2.0 mL/kg body weight to ensure accurate measurements 2

Gas Volume Standardization

Gas volumes must be expressed in STPD (standard temperature, pressure, dry) conditions for VCO2/VE calculations 2

Normal Values and Clinical Interpretation

  • Normal VD/VT at rest: Approximately 0.34 (or less than 0.35) 2
  • Normal VD/VT during exercise: Decreases to ≤0.10 in healthy individuals 2
  • Abnormal values: In mechanically ventilated patients with ARDS, VD/VT >0.6 is clearly abnormal and associated with significantly higher mortality 2

Physiological Meaning

The VD/VT ratio represents the fraction of each breath that is "wasted" on ventilating both the anatomic dead space (conducting airways) and alveolar dead space (ventilated but unperfused or underperfused alveoli) 1

An increase in VD/VT reflects increased inefficiency of ventilation due to ventilation-perfusion mismatching or right-to-left shunt, requiring increased minute ventilation to maintain PaCO2 1

Common Pitfalls

Breathing pattern significantly affects VD/VT measurements: Rapid shallow breathing increases VD/VT even without true ventilation-perfusion abnormalities 1

Correction for mechanical dead space: When alveolar dead space is present, less than the full volume of attached mechanical dead space appears in the measured dead space, and traditional subtraction of mechanical dead space leads to underestimation of physiological dead space 3

References

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Guideline

Computing Deadspace Fraction in Mechanically Ventilated Adults

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2026

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.