How FEV1 is Calculated
FEV1 is the maximum volume of air exhaled during the first second of a forced expiration from full inspiration, measured in liters at body temperature and pressure saturated with water vapor (BTPS), with timing initiated from time zero defined by back-extrapolation. 1
Measurement Procedure
Patient Preparation and Maneuver Execution
- The patient inhales to total lung capacity (full inspiration) 1
- The patient then exhales as forcefully and completely as possible into the spirometer 2
- At least three acceptable forced expiratory curves must be obtained 2, 1
Determining Time Zero
- Time zero is established by back-extrapolation from the steepest part of the volume-time curve, not from when the patient begins exhaling 1
- This back-extrapolation method is the gold standard, as alternative methods using flow or volume thresholds show too much variability to provide uniformly accurate equivalents 3
Selecting the Final FEV1 Value
- Record the largest FEV1 value from all usable curves, even if it does not come from the same curve as the largest FVC 2, 1
- This approach differs from selecting a single "best" curve and ensures you capture the patient's true maximum performance 2
Quality Control Criteria
Within-Maneuver Acceptability
Each spirogram must be free from artifacts including: 2, 1
- Cough during the first second of exhalation
- Glottic closure affecting measurement
- Early termination or submaximal effort
- Leaks or obstructed mouthpiece
The maneuver must have a good start with extrapolated volume <5% of FVC or <0.15 L, whichever is greater 2, 1
Between-Maneuver Repeatability
- The two largest FEV1 values must be within 0.150 L of each other 2, 1
- The two largest FVC values must also be within 0.150 L of each other 2
- If these criteria are not met after three acceptable maneuvers, continue testing up to a maximum of eight attempts 2
Technical Considerations
BTPS Correction
- All measurements must be corrected to BTPS conditions 2
- The BTPS correction factor can be as large as 10% depending on environmental temperature 2
- Spirometer temperature should be measured for each maneuver, not assumed constant, as errors up to 6% in FEV1 can occur if ambient temperature is used instead of actual spirometer temperature 2
Equipment Accuracy Requirements
- Spirometers must have accuracy of ±3% of reading or ±0.050 L (whichever is greater) for FEV1 measurement 2
- Flow range must accommodate 0-14 L/s 2
- Resistance must be <1.5 cmH₂O·L⁻¹·s⁻¹ 2
Common Pitfalls to Avoid
- Do not reject results solely based on poor repeatability—label them appropriately and leave interpretation to the clinician's discretion 2
- Do not use maneuvers with unacceptable starts or coughs during the first second when determining the best FEV1 value 2
- Maneuvers with early termination may still be used for selecting the largest FEV1, as long as the first second is artifact-free 2
- If a patient shows progressive reduction in FEV1 exceeding 20% from start value across sequential maneuvers, terminate testing for patient safety 2