Visual Inspection and Laboratory Assessment
The most straightforward initial assessment for hemolysis is visual inspection of plasma or serum for red or pink discoloration, followed by quantitative measurement using the hemolysis index (HI) on automated chemistry analyzers or spectrophotometric methods. 1
Visual Assessment
Red or pink tingeing of plasma or serum indicates significant hemolysis has occurred, which may render samples less useful for many biomarker studies 1. This simple visual check should be performed immediately after centrifugation and before freezing samples. Samples that appear red, orange, or contain visible discoloration should be flagged or discarded depending on the testing requirements 2.
Quantitative Laboratory Methods
Hemolysis Index (HI)
Modern clinical chemistry analyzers automatically measure the hemolysis index spectrophotometrically during sample processing. This provides an objective, quantitative assessment of free hemoglobin in the sample. The HI is widely available and should be routinely reported 3.
Specific Biochemical Markers
When assessing for in vivo hemolysis (actual red blood cell destruction in the patient), measure:
- Lactate dehydrogenase (LDH): The most specific biochemical indicator of hemolysis. Elevation >2.5 times the upper limit of normal is highly suggestive 4
- Plasma free hemoglobin: Values >40 mg/dL indicate significant hemolysis 4
- Haptoglobin: Decreased levels signal hemolysis, though this can be less sensitive 4
- Unconjugated bilirubin: Elevated in hemolysis 4
- Hemoglobin/hematocrit: Decreased levels 4
For diagnostic confirmation of hemolysis, simultaneously elevated LDH and plasma free hemoglobin together are diagnostic 4.
Point-of-Care Testing
Novel POCT devices like HemCheck™ can detect hemolysis directly on whole blood samples before centrifugation by measuring free hemoglobin levels 5, 6. These devices use a cartridge system with vertical and lateral flow filtration, with a red LED indicator illuminating when hemolysis is detected. This is particularly valuable in emergency departments and ICUs where immediate assessment is needed.
Critical Preanalytical Considerations
Sample Collection Factors That Cause Hemolysis:
- Use of butterfly needles (prefer straight needles) 7
- Tourniquet pressure maintained too long 7
- Drawing blood into syringes before transferring to tubes 7
- Residual alcohol on skin not allowed to evaporate 1
- Vigorous mixing or shaking of tubes
- Small gauge needles with rapid aspiration
Sample Handling:
- Process samples within 1-2 hours of collection 7
- Keep at room temperature, never on ice (causes platelet activation) 2
- Use proper centrifugation speeds (2000g for 10 minutes) 1
- Avoid hemolysis from freeze-thaw cycles 1
Common Pitfalls
Do not confuse in vitro hemolysis (specimen handling artifact) with in vivo hemolysis (actual patient condition). In vitro hemolysis affects potassium results most dramatically, potentially causing false hyperkalemia that could lead to inappropriate treatment 6. Up to 12-14% of emergency department blood gas samples may be hemolyzed without detection 6.
LDH can be elevated from non-hemolytic sources including liver, lungs, and muscle tissue. When LDH is elevated, confirm hemolysis by checking plasma free hemoglobin simultaneously 4.
Reporting Standards
Document and report 8:
- Presence and degree of hemolysis (slight, mild, moderate, heavy)
- Hemolysis index values when available
- Whether samples were discarded or results reported with caveats
- Collection and processing details that may have contributed to hemolysis