Through-and-Through Wire Selection for Graft Deployment

For through-and-through (T&T) technique graft deployment, use a stiff 0.035-inch guidewire, specifically an extra-stiff or Lunderquist-type wire, to provide maximum support for endograft delivery through tortuous anatomy.

Wire Specifications

The evidence consistently supports specific wire characteristics for T&T techniques:

  • Wire diameter: 0.035-inch is the standard 1, 2, 3
  • Wire type: Extra-stiff or Lunderquist guidewires provide optimal support 4, 2
  • Wire characteristics: Stiff-bodied construction is essential for maintaining tension and straightening tortuous anatomy 1, 3

Technical Rationale

The T&T technique requires establishing wire continuity between two access points (commonly femoral-to-brachial, femoral-to-transapical, or femoral-to-carotid). The primary purpose is to create a rigid rail that straightens vascular tortuosity and provides sufficient support to advance large-bore delivery systems that would otherwise fail with standard single-access techniques 3.

When the stiff 0.035-inch wire is externalized and tightened from both ends, it effectively:

  • Straightens severe curves in the aortoiliac system
  • Provides a stable platform for large sheath advancement
  • Enables delivery of endografts to challenging proximal landing zones 1, 2

Access Point Considerations

While the wire specification remains constant, access combinations vary by anatomy:

  • Femoral-to-brachial: Most common for thoracic arch procedures 3
  • Femoral-to-transapical: Used when valve crossing is acceptable or in ruptured scenarios 2
  • Femoral-to-carotid: Alternative for arch repair when transapical access is not feasible 5

Initial Wire Placement Strategy

Before establishing the T&T wire, initial crossing may require a different approach. A hydrophilic-coated 0.035-inch guidewire or smaller 0.014-inch wire can be used for initial navigation through tortuous or aneurysmal segments 1. Once position is achieved, exchange to the stiff 0.035-inch wire through a support catheter (such as a 5F vertebral catheter) before externalization 1.

Critical Technical Points

Avoid using smaller diameter wires (0.018-inch, 0.021-inch, or 0.025-inch) for the final T&T configuration, as these lack the rigidity needed for large-bore device delivery and are prone to kinking 6. While a 0.025-inch wire may be mentioned in biliary procedures 6, this does not translate to aortic endograft deployment where significantly greater support is required.

The 1.5-mm-radius J-tip 0.035-inch wire recommended for transradial access 7 is appropriate for peripheral navigation but insufficient for T&T graft deployment—you need the extra-stiff variant, not the standard J-tip configuration.

Common Pitfall

Do not confuse the 0.035-inch wires used for initial TAVR access 8, 9 with T&T technique requirements. TAVR procedures describe exchanging to "stiffer" 0.035-inch wires (super or extra-stiff) after initial crossing 8, 9, which aligns with T&T principles, but the T&T technique specifically requires maintaining that stiff wire configuration between two access points with external traction—a fundamentally different mechanical setup than single-access procedures.

References

Research

Transapical to Femoral Through-and-Through Wire Technique for Ruptured Aortic Arch Aneurysm Management: Case Report and Review of Literature.

Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists, 2025

Research

Ascending Aorta Nose-Cone Loop Technique as Bail Out for Precise Branched Endovascular Aortic Arch Endograft Delivery Without Valve Re-Crossing.

Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists, 2025

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.