Through-and-Through Wire Selection for Graft Deployment
For through-and-through (T&T) technique graft deployment, use a stiff 0.035-inch guidewire, specifically an extra-stiff or Lunderquist-type wire, to provide maximum support for endograft delivery through tortuous anatomy.
Wire Specifications
The evidence consistently supports specific wire characteristics for T&T techniques:
- Wire diameter: 0.035-inch is the standard 1, 2, 3
- Wire type: Extra-stiff or Lunderquist guidewires provide optimal support 4, 2
- Wire characteristics: Stiff-bodied construction is essential for maintaining tension and straightening tortuous anatomy 1, 3
Technical Rationale
The T&T technique requires establishing wire continuity between two access points (commonly femoral-to-brachial, femoral-to-transapical, or femoral-to-carotid). The primary purpose is to create a rigid rail that straightens vascular tortuosity and provides sufficient support to advance large-bore delivery systems that would otherwise fail with standard single-access techniques 3.
When the stiff 0.035-inch wire is externalized and tightened from both ends, it effectively:
- Straightens severe curves in the aortoiliac system
- Provides a stable platform for large sheath advancement
- Enables delivery of endografts to challenging proximal landing zones 1, 2
Access Point Considerations
While the wire specification remains constant, access combinations vary by anatomy:
- Femoral-to-brachial: Most common for thoracic arch procedures 3
- Femoral-to-transapical: Used when valve crossing is acceptable or in ruptured scenarios 2
- Femoral-to-carotid: Alternative for arch repair when transapical access is not feasible 5
Initial Wire Placement Strategy
Before establishing the T&T wire, initial crossing may require a different approach. A hydrophilic-coated 0.035-inch guidewire or smaller 0.014-inch wire can be used for initial navigation through tortuous or aneurysmal segments 1. Once position is achieved, exchange to the stiff 0.035-inch wire through a support catheter (such as a 5F vertebral catheter) before externalization 1.
Critical Technical Points
Avoid using smaller diameter wires (0.018-inch, 0.021-inch, or 0.025-inch) for the final T&T configuration, as these lack the rigidity needed for large-bore device delivery and are prone to kinking 6. While a 0.025-inch wire may be mentioned in biliary procedures 6, this does not translate to aortic endograft deployment where significantly greater support is required.
The 1.5-mm-radius J-tip 0.035-inch wire recommended for transradial access 7 is appropriate for peripheral navigation but insufficient for T&T graft deployment—you need the extra-stiff variant, not the standard J-tip configuration.
Common Pitfall
Do not confuse the 0.035-inch wires used for initial TAVR access 8, 9 with T&T technique requirements. TAVR procedures describe exchanging to "stiffer" 0.035-inch wires (super or extra-stiff) after initial crossing 8, 9, which aligns with T&T principles, but the T&T technique specifically requires maintaining that stiff wire configuration between two access points with external traction—a fundamentally different mechanical setup than single-access procedures.