What to Do If a Patient Coughs or Sneeze During a PET/CT Scan
If a patient coughs or sneezes during a PET/CT scan, the affected portion of the scan should be repeated to avoid motion artifacts that can compromise image quality and diagnostic accuracy.
Understanding the Problem
Coughing or sneezing during PET/CT acquisition creates significant motion artifacts that are more severe than normal respiratory motion. These sudden movements cause:
- Misalignment between PET and CT images, leading to incorrect attenuation correction
- Blurring of lesions, which reduces SUV accuracy and can miss or mischaracterize disease
- Curvilinear artifacts that can mimic or obscure pathology
The EANM guidelines emphasize that images should be reviewed before the patient leaves the department to ensure that the examination is technically satisfactory and can properly address the clinical question 1. This is the critical checkpoint where motion artifacts from coughing/sneezing must be identified.
Immediate Management Protocol
During the Scan:
- Pause the acquisition immediately if you observe the patient cough or sneeze
- Note the bed position and timing of the motion event
- Allow the patient to settle and resume normal shallow breathing
After Scan Completion:
- Review images at the console before dismissing the patient 1
- Identify affected bed positions - look for:
- Obvious misregistration between PET and CT
- Curvilinear artifacts at the diaphragm or organ boundaries
- Unusual SUV patterns or blurred lesions
- Repeat only the affected bed position(s) rather than the entire scan
Technical Considerations
The standard EANM protocol specifies that CT should be performed during tidal or shallow breathing 1. A cough or sneeze violates this fundamental assumption and creates motion far exceeding normal respiratory excursion. Research demonstrates that even normal respiratory motion causes significant artifacts 2, 3, 4, so the violent motion from coughing/sneezing will invariably compromise image quality.
Key Points:
- Normal respiratory motion already causes measurable misalignment (mean 3.5 mm) 2
- Coughing/sneezing produces far greater displacement, making artifacts inevitable
- SUV measurements can be underestimated by >20% with motion artifacts 2
- Diagnostic accuracy suffers particularly for small lesions (<3 cm) 2
Prevention Strategies
Patient preparation is essential:
- Instruct patients to suppress coughs if possible during the brief scan time (2-5 minutes per bed position)
- Provide cough suppressants if the patient has a known cough
- Allow patients to clear their throat before starting each bed position
- Use shorter acquisition times per bed position when feasible to minimize the window for motion events
Common Pitfalls to Avoid
- Don't assume the scan is adequate without review - motion artifacts may not be obvious on quick inspection
- Don't dismiss the patient before image review - recalling them for a repeat scan is logistically difficult and may require rescheduling
- Don't repeat the entire scan unnecessarily - only affected bed positions need repeating, minimizing radiation exposure and scan time
- Don't ignore subtle misregistration - even small artifacts can affect quantitative measurements and treatment planning
The guideline principle is clear: ensure technical adequacy before patient departure 1. This applies directly to motion events like coughing or sneezing that compromise image quality.