Can a depleted pacemaker battery that causes mode shift lead to pacemaker syndrome?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: July 14, 2025View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

Depleted Pacemaker Battery Mode Shift Can Cause Pacemaker Syndrome

Yes, a depleted pacemaker battery that causes mode shift can definitely cause pacemaker syndrome. When a pacemaker battery reaches depletion, many devices automatically revert to a backup pacing mode (typically VVI), which can disrupt atrioventricular (AV) synchrony and lead to pacemaker syndrome 1.

Understanding Pacemaker Syndrome

Pacemaker syndrome is defined as the occurrence of overt symptoms due to adverse hemodynamics resulting from:

  • Loss of AV synchrony
  • Occurrence of ventriculoatrial conduction
  • Atrial contraction against closed AV valves 1

Common symptoms include:

  • Fatigue
  • Chest discomfort
  • Dyspnea
  • Cough
  • Confusion
  • Presyncope or syncope 1

How Battery Depletion Causes Pacemaker Syndrome

When a pacemaker battery reaches its elective replacement indicator (ERI) or end-of-life:

  1. The device often automatically switches from its programmed mode to a backup mode (typically VVI - ventricular pacing inhibited by sensed ventricular activity) 2
  2. This mode shift from a dual-chamber mode (like DDD) to VVI eliminates AV synchrony
  3. Loss of AV synchrony can lead to:
    • Decreased cardiac output
    • Venous "cannon A waves"
    • Sudden increase in atrial pressure
    • Systemic hypotensive reflex response 3

Clinical Significance

This phenomenon is particularly problematic in:

  • Patients who were initially implanted with dual-chamber pacemakers due to risk of pacemaker syndrome with ventricular pacing
  • Patients with intact retrograde ventriculoatrial conduction
  • Patients with heart failure who benefit from optimized AV timing 1

The risk is heightened in patients with:

  • Sinus node dysfunction with normal AV conduction
  • Patients who rely on atrial contribution to cardiac output 1

Management Approach

When a pacemaker battery is depleted and mode shift occurs:

  1. Urgent replacement of the pulse generator is indicated if pacemaker syndrome symptoms develop

  2. Monitor for symptoms including:

    • Hypotension
    • Dizziness/lightheadedness
    • Fatigue
    • Dyspnea
    • Palpitations
    • Cannon waves in jugular venous pulse
  3. Temporary programming may provide relief until replacement:

    • If possible, program to minimize ventricular pacing
    • Adjust rate response settings to minimize symptoms

Important Considerations

  • Older pacemakers (pre-2005) are more likely to experience problematic mode shifts during battery depletion 4
  • The severity of pacemaker syndrome varies widely among patients
  • Some patients may be asymptomatic despite loss of AV synchrony
  • Pacemaker syndrome can occur in up to 83% of patients paced in VVI mode who were previously in a physiologic pacing mode 1

Prevention

  • Regular follow-up to monitor battery status
  • Replacement before reaching ERI in patients at high risk for pacemaker syndrome
  • Understanding the specific behavior of the implanted device at battery depletion
  • For patients undergoing device upgrades (e.g., to CRT), be aware that a depleted old device may revert to VVI mode and interfere with the new system 2

Pacemaker syndrome remains an important clinical entity that can significantly impact morbidity and quality of life, particularly when caused by unexpected mode shifts due to battery depletion.

References

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Research

A devices' game of thrones: cardiac resynchronization therapy vs. pacemaker.

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2017

Research

The pacemaker syndrome: old and new causes.

Clinical cardiology, 1991

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.