Pacing Spike Early After Paced QRS: Failure to Sense
Yes, a pacing spike occurring early after a paced QRS complex typically indicates failure to sense, which means the pacemaker is not detecting the intrinsic or paced cardiac activity and is inappropriately delivering additional pacing stimuli.
Understanding Sensing Failure
Sensing failure occurs when the pacemaker fails to detect intrinsic or previously paced cardiac electrical activity. This results in the pacemaker delivering unnecessary pacing stimuli, which can be seen as pacing spikes occurring soon after a QRS complex.
Mechanisms of Sensing Failure:
Intraventricular Conduction Defects
- Delayed arrival of ventricular depolarization at the pacemaker sensing electrodes due to intraventricular conduction disturbances 1
- Particularly common in patients with right bundle branch block (RBBB) during conducted sinus beats or with ectopic beats arising from the left ventricle
- Activation of the right ventricular apex can be delayed up to 65 msec in sinus beats with RBBB 1
Technical Factors
- Lead displacement or dislodgement
- Insulation defects
- Lead fracture
- Inappropriate sensitivity settings
- Battery depletion
Clinical Implications
Sensing failure has several important clinical implications:
Proarrhythmic Potential
- Pacing spikes falling on vulnerable periods of the cardiac cycle (T-wave) can potentially trigger ventricular arrhythmias
- Increased risk of sudden cardiac death in paced patients (12-31%) 2
Hemodynamic Consequences
- Disruption of optimal AV synchrony
- Potential for pacemaker syndrome
- Reduced cardiac output
Battery Depletion
- Unnecessary pacing shortens battery life
Diagnosis
When a pacing spike is observed early after a paced QRS:
ECG Monitoring
Pacemaker Interrogation
- Essential to confirm sensing failure 3
- Check sensing thresholds, lead impedance, and battery status
Evaluate for Lead Displacement
Management
Management depends on the patient's pacemaker dependency:
For Pacemaker-Dependent Patients
- Urgent lead repositioning if lead dislodgement is confirmed 3
- Temporary pacing may be necessary while awaiting definitive management
For Non-Pacemaker-Dependent Patients
- Management decisions should be based on symptoms and underlying rhythm 3
- Reprogramming of sensing parameters may resolve the issue
Specific Interventions
- Adjust sensitivity settings
- Replace leads if damaged
- Reposition displaced leads
- Replace pulse generator if battery depletion is the cause
Prevention
To minimize sensing failures:
During Implantation
- Proper lead positioning
- Adequate fixation
- Verification of appropriate sensing thresholds
During Generator Replacement
- Create an appropriately sized pocket for the generator
- Secure generator with strong sutures
- Careful handling of leads
- Fix sleeves to appropriate tightness 3
Patient Education
- Post-procedure activity limitations
- Recognition of symptoms that may indicate pacemaker malfunction
Special Considerations
QRS Duration
Mode Selection
- In patients with intraventricular conduction defects, mode selection should consider potential sensing issues
- Asynchronous pacing modes (AOO, VOO, DOO) may be considered temporarily when sensing issues are problematic
Remember that sensing failure should be distinguished from other pacemaker issues such as failure to capture or inappropriate inhibition of pacing, which may present differently on ECG.