What are the recommended settings for inspiratory trigger (in L/min), expiratory trigger (as a percentage of peak inspiratory flow), PS (Pressure Support) rise time (in seconds), and bias flow for weaning a 1-month-old child from Synchronized Intermittent Mandatory Ventilation (SIMV) volume control?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: October 7, 2025View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

Ventilator Settings for Weaning a 1-Month-Old Child from SIMV Volume Control

For weaning a 1-month-old from SIMV volume control ventilation, set the inspiratory trigger at 0.5-0.6 L/min, expiratory trigger at 15-25% of peak inspiratory flow, PS rise time at 0.1-0.2 seconds, and bias flow at 4-6 L/min to optimize patient-ventilator synchrony and facilitate successful extubation. 1

Inspiratory Trigger (Flow Trigger)

  • Set the inspiratory trigger sensitivity at 0.5-0.6 L/min for infants to optimize patient-ventilator synchrony while avoiding auto-triggering 1, 2
  • The trigger should be sensitive enough to detect the infant's inspiratory effort but not so sensitive that it auto-triggers from circuit movement or cardiac oscillations 1
  • Measure flow near the Y-piece of the patient circuit for children <10 kg to improve trigger sensitivity 1

Expiratory Trigger (Cycling)

  • Set the expiratory trigger at 15-25% of peak inspiratory flow for infants to prevent premature cycling and optimize synchrony 1, 3
  • Monitor the flow-time scalar to ensure proper cycling from inspiration to expiration 1
  • Adjust expiratory trigger based on observation of flow-time waveforms to prevent air-trapping, especially important in this age group 1

Pressure Support Rise Time

  • Set PS rise time at 0.1-0.2 seconds for infants to provide smooth pressure delivery without causing flow starvation 1, 3
  • Too fast rise time may cause pressure overshooting and patient discomfort 1
  • Too slow rise time may increase work of breathing and lead to patient-ventilator asynchrony 2

Bias Flow

  • Set bias flow at 4-6 L/min for infants to ensure adequate flow availability for triggering without increasing work of breathing 1
  • Higher bias flow may be needed if there is increased respiratory drive during weaning 3
  • Excessive bias flow can increase dead space ventilation and work of breathing 1

Weaning Strategy

  • Target patient-ventilator synchrony as a primary goal during weaning 1
  • Start weaning as soon as possible and perform daily extubation readiness testing 1
  • When using SIMV for weaning, gradually reduce the mandatory rate while maintaining adequate spontaneous breathing support 4, 2
  • Consider adding pressure support to SIMV to reduce work of breathing during spontaneous breaths 2, 5
  • Monitor pressure-time and flow-time scalars to assess patient-ventilator interaction 1

Monitoring Parameters During Weaning

  • Measure SpO2 continuously and maintain ≥95% in healthy lungs 1
  • Monitor end-tidal CO2 in all ventilated children 1
  • Target PCO2 of 35-45 mmHg for healthy lungs, with higher values acceptable during weaning if pH >7.20 1
  • Observe for signs of increased work of breathing, which may indicate inappropriate ventilator settings 2, 6
  • Monitor respiratory rate during weaning; acceptable ranges for a 1-month-old are typically 20-50 breaths per minute 6

Common Pitfalls to Avoid

  • Setting inspiratory trigger too sensitive (causing auto-triggering) or too insensitive (increasing work of breathing) 1, 2
  • Inappropriate expiratory trigger settings leading to air-trapping or premature cycling 1
  • Inadequate pressure support during spontaneous breaths, increasing work of breathing 2, 5
  • Weaning too rapidly, which may lead to fatigue and extubation failure 1
  • Failure to monitor for patient-ventilator asynchrony, which increases work of breathing 1

By optimizing these ventilator settings during weaning, you can reduce the work of breathing, promote patient-ventilator synchrony, and increase the likelihood of successful extubation in your 1-month-old patient.

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.