How Pressure Support Ventilation Works
Pressure support ventilation (PSV) is a patient-triggered, pressure-limited, and flow-cycled mode of ventilation that augments the patient's spontaneous inspiratory efforts with a clinician-selected level of positive airway pressure to reduce work of breathing. 1, 2
Basic Mechanism of Action
- In PSV, the patient's respiratory effort triggers the ventilator to deliver a preset pressure during inspiration, while the patient controls both respiratory rate and timing of each breath 1
- The ventilator delivers a constant pressure throughout inspiration until the flow decreases to a certain threshold (typically 25% of peak flow), at which point the ventilator cycles to expiration 2
- If the patient fails to make respiratory efforts, no respiratory assistance will occur, unless the ventilator has a backup rate feature (many modern ventilators incorporate a backup rate of 6-8 breaths per minute) 1
Key Components and Settings
- Trigger: Patient's inspiratory effort initiates breath delivery (flow or pressure triggered) 2
- Pressure Support Level: Clinician-selected pressure that determines the amount of assistance provided during inspiration 3
- Rise Time (Pressurization Rate): Speed at which the ventilator reaches the set pressure support level after triggering 2
- Flow Cycling Criteria: The percentage of peak inspiratory flow at which the ventilator cycles from inspiration to expiration (typically 25%, but adjustable on newer ventilators) 2
Physiological Effects
- PSV reduces respiratory muscle workload by decreasing the work of breathing required by the patient 3
- It improves patient comfort compared to volume-controlled ventilation modes 4
- PSV allows for a more balanced pressure and volume change form of muscle work for the patient 3
- The delivered tidal volume depends on the pressure support level, patient effort, and respiratory system mechanics (resistance and compliance) 2
Clinical Applications
Advantages
- PSV is more comfortable for patients compared to volume-controlled continuous mandatory ventilation (VC-CMV), with studies showing significantly higher comfort scores 4
- It can be used during weaning from mechanical ventilation, as it allows patients to gradually resume respiratory muscle work 1
- PSV can be used in spontaneous breathing trials (SBTs) with modest pressure support (5-8 cmH2O) to overcome the resistance of the endotracheal tube 1
Potential Issues
- High levels of PSV may cause sleep disruption from periodic breathing due to hyperventilation and hypocapnia 1
- If pressure support delivers higher-than-needed alveolar minute ventilation, it can lead to central apneas during sleep 1
- In patients with increased ventilatory demand, fixed PSV may not adequately respond to changing patient needs 5
Advanced PSV Variations
Volume-Assured Pressure Support (VAPS): Automatically adjusts inspiratory pressure between set minimum and maximum limits to deliver a targeted tidal volume 6
Volume Support Ventilation (VSV): Continuously adjusts pressure support level to deliver a preset tidal volume 5
- Caution: When ventilatory demand increases, VSV may decrease the pressure support provided, potentially causing respiratory distress 5
Practical Considerations
- Triggering difficulties during PSV are usually due to intrinsic positive end-expiratory pressure (auto-PEEP) 2
- Patient-ventilator dyssynchrony may occur if the flow at which the ventilator cycles to exhalation does not coincide with the termination of neural inspiration 2
- Adjusting the rise time and flow-termination criteria can improve patient-ventilator synchrony 2
- Ventilator waveforms are useful for appropriately adjusting PSV settings 2
Common Pitfalls
- Setting excessive pressure support levels can cause hyperventilation, leading to central apneas during sleep 1
- Inadequate pressure support may result in increased work of breathing and patient discomfort 3
- Failure to set a backup rate in patients with central hypoventilation or apneas can lead to inadequate ventilation 6
- Patient-ventilator asynchrony can occur if flow cycling criteria are not properly adjusted 2