Measuring Peak Expiratory Flow Rate in the ICU
In mechanically ventilated ICU patients, peak expiratory flow rate (PEF) can be measured directly using the ventilator's built-in flow meter without disconnecting the patient, or alternatively with a portable electronic peak flow meter connected to the endotracheal tube. 1
Methods for PEF Measurement in Ventilated ICU Patients
Using the Ventilator's Built-In Flow Meter (Preferred Method)
- Modern ICU ventilators incorporate flow velocity measurement capabilities that allow direct assessment of cough peak expiratory flow without patient disconnection. 1
- This method is feasible in routine practice with cooperative mechanically ventilated patients who are in the weaning process. 1
- The technique works best in patients ventilated with pressure support <15 cm H₂O and PEEP <9 cm H₂O. 1
- Ventilator-measured CPF values correlate moderately well (r=0.63) with portable peak flow meter measurements, though ventilator readings are typically higher (mean 72.6 L/min vs 31.1 L/min). 1
Using a Portable Electronic Peak Flow Meter
- An electronic portable handheld peak flow meter can be connected directly to the endotracheal tube for measurement. 1
- This requires temporary disconnection from the ventilator circuit, which may not be ideal for unstable patients. 1
- The portable device provides lower absolute values but maintains correlation with ventilator measurements. 1
Measurement Technique for Ventilated Patients
Patient Preparation
- The patient must be cooperative and able to follow commands to generate a maximal cough effort. 1
- Ensure the patient is in a semi-recumbent or upright position if tolerated. 2
- The measurement is typically performed on the day of planned extubation when the patient is most alert and cooperative. 1
Execution of Measurement
- Instruct the patient to take a deep breath to total lung capacity, then cough as hard and fast as possible. 2, 1
- The exhalation should be forceful and explosive rather than prolonged. 2
- Record the highest value from three attempts. 2
- Results are expressed in liters per minute (L/min). 2, 3
Clinical Interpretation in the ICU Context
Threshold Values
- A ventilator CPF with area under the ROC curve of 0.84 can predict a portable meter CPF <35 L/min, which is a commonly used threshold for extubation readiness. 1
- However, neither ventilator CPF nor portable meter CPF reliably predicted re-intubation within 72 hours in one study. 1
Important Caveats
- The ventilator-measured values are systematically higher than portable meter values due to differences in measurement technique and resistance characteristics. 1
- PEF is highly effort-dependent and requires proper patient cooperation for accurate results. 2
- The measurement may be less reliable in patients with altered mental status, heavy sedation, or neuromuscular weakness. 4
Equipment Considerations
- Ensure the ventilator flow meter is properly calibrated according to manufacturer specifications. 4
- If using a portable device, use disposable mouthpieces (or in this case, ensure sterile connection to the endotracheal tube) to prevent cross-contamination. 2
- The resistance of the measurement device should not exceed American Thoracic Society recommendations, particularly at flow rates >600 L/min. 5
Alternative Approach for Non-Ventilated ICU Patients
For ICU patients who are not mechanically ventilated:
- Use standard peak flow meter technique with the patient in sitting or standing position if possible. 2
- The patient should take a deep breath to maximum inspiration, seal lips tightly around the mouthpiece, and blow out as hard and fast as possible. 2
- Ensure the peak flow meter indicator is at zero before starting. 2
- Record the highest of three attempts. 2
Limitations in the ICU Setting
- Maximum inspiratory pressure measurements have poor reproducibility in critically ill patients and are of limited use for decision-making. 4
- Traditional predictors like vital capacity and minute ventilation are frequently falsely positive or negative for weaning outcomes. 4
- The ratio of respiratory frequency to tidal volume remains the most reliable simple predictor of weaning outcome, more so than isolated PEF measurements. 4