When to Adjust Blood Flow During Hemodialysis
Blood flow rate should be decreased to 50-100 mL/min for 15 seconds at the end of dialysis specifically for accurate post-dialysis BUN sampling, then either maintained at this slow flow or stopped completely depending on your sampling technique. 1
Primary Indication: Post-Dialysis Blood Sampling
The most clearly defined situation requiring blood flow adjustment occurs at treatment completion for laboratory sampling 1:
Slow Flow/Stop Pump Sampling Technique
At dialysis completion:
- Turn off dialysate flow and reduce ultrafiltration to 50 mL/h or minimum setting 1
- Decrease blood flow to 50-100 mL/min for 15 seconds to clear the arterial line dead space of any recirculated blood 1
- You may need to manually adjust venous pressure limits downward to prevent pump shut-off 1
Purpose: This fills the arterial needle tubing with non-recirculated blood, eliminating access recirculation artifact that would falsely lower the BUN and artificially inflate Kt/V calculations 1
Then proceed with either:
- Slow flow technique: Maintain blood pump at 50-100 mL/min and draw sample from arterial port 1
- Stop pump technique: Immediately stop the pump, clamp lines, and draw sample 1
Critical timing: Sampling at 0.25-0.50 minutes post-dialysis (corresponding to this slow flow/stop pump technique) provides the most accurate measurement for single-pool urea kinetic modeling, before significant urea rebound occurs 1
Situations Requiring Blood Flow Optimization During Treatment
Inadequate Vascular Access Flow
When access flow is compromised (<600 mL/min), you should still increase blood flow rate to maximize Kt/V 2:
- Even with low access flow rates, increasing dialyzer blood flow from 200 to 300 to 400 mL/min generally increases delivered Kt/V 2
- Access recirculation remains minor (0.9% ± 0.6%) even at higher blood flow rates in patients with low access flow 2
- The primary solution for inadequate blood flow is extending treatment time, not reducing blood flow 3
Key threshold: Blood flow <300 mL/min indicates catheter dysfunction and requires intervention 4:
- Prepump arterial pressure more negative than -250 mmHg signals inadequate flow 4
- For high-efficiency dialysis, blood flow >300 mL/min is required to achieve target spKt/V of 1.2 4
- 15% of catheter treatments have blood flow <300 mL/min, leading to underdialysis 4
Balancing Blood Flow with Dialysate Flow
Maintain a blood flow to dialysate flow ratio of 1:2 for optimal efficiency 5:
- If dialysate flow must be reduced (e.g., from 500 to 400 mL/min), calculate the appropriate blood flow using the K₀A equation to maintain equivalent dialysis efficiency 5
- Example: When reducing dialysate flow from 500 to 400 mL/min with blood flow of 200 mL/min, increase blood flow to 210 mL/min to maintain clearance 5
Catheter-Specific Considerations
Central venous catheters may require blood flow adjustment based on position and design 4:
- Femoral catheters have significantly greater recirculation than internal jugular catheters (13.1% vs 0.4%) 4
- Femoral catheters <20 cm have recirculation of 26.3% vs 8.3% for those >20 cm 4
- Newer catheter designs can achieve 400 mL/min or greater when properly placed 4
- Exception: Pediatric or smaller adult catheters are not designed for flows >300 mL/min 4
Common Pitfalls to Avoid
Do not sample immediately at time zero (blood pump still running at full speed): This causes major BUN reduction from recirculated blood still in the access, falsely elevating Kt/V 1
Do not assume that reducing blood flow will improve Kt/V in patients with low access flow: Studies demonstrate that even with compromised access, higher blood flow rates generally increase delivered dose 2
Do not reduce blood flow as the primary solution for inadequate dialysis: The prescribed dose should be Kt/V 1.3 (URR ~70%) to ensure minimum delivered dose of 1.2 is achieved 1. When blood flow is limited, extend treatment time instead 1, 3
Monitor for factors that compromise effective blood flow 1:
- Access recirculation reducing concentration gradient
- Blood pump/dialysate flow calibration errors
- Dialyzer clotting during treatment
- High pre-pump negative pressure causing tubing collapse