When should blood flow be adjusted during hemodialysis (HD)?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: December 20, 2025View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

When to Adjust Blood Flow During Hemodialysis

Blood flow rate should be decreased to 50-100 mL/min for 15 seconds at the end of dialysis specifically for accurate post-dialysis BUN sampling, then either maintained at this slow flow or stopped completely depending on your sampling technique. 1

Primary Indication: Post-Dialysis Blood Sampling

The most clearly defined situation requiring blood flow adjustment occurs at treatment completion for laboratory sampling 1:

Slow Flow/Stop Pump Sampling Technique

At dialysis completion:

  • Turn off dialysate flow and reduce ultrafiltration to 50 mL/h or minimum setting 1
  • Decrease blood flow to 50-100 mL/min for 15 seconds to clear the arterial line dead space of any recirculated blood 1
  • You may need to manually adjust venous pressure limits downward to prevent pump shut-off 1

Purpose: This fills the arterial needle tubing with non-recirculated blood, eliminating access recirculation artifact that would falsely lower the BUN and artificially inflate Kt/V calculations 1

Then proceed with either:

  • Slow flow technique: Maintain blood pump at 50-100 mL/min and draw sample from arterial port 1
  • Stop pump technique: Immediately stop the pump, clamp lines, and draw sample 1

Critical timing: Sampling at 0.25-0.50 minutes post-dialysis (corresponding to this slow flow/stop pump technique) provides the most accurate measurement for single-pool urea kinetic modeling, before significant urea rebound occurs 1

Situations Requiring Blood Flow Optimization During Treatment

Inadequate Vascular Access Flow

When access flow is compromised (<600 mL/min), you should still increase blood flow rate to maximize Kt/V 2:

  • Even with low access flow rates, increasing dialyzer blood flow from 200 to 300 to 400 mL/min generally increases delivered Kt/V 2
  • Access recirculation remains minor (0.9% ± 0.6%) even at higher blood flow rates in patients with low access flow 2
  • The primary solution for inadequate blood flow is extending treatment time, not reducing blood flow 3

Key threshold: Blood flow <300 mL/min indicates catheter dysfunction and requires intervention 4:

  • Prepump arterial pressure more negative than -250 mmHg signals inadequate flow 4
  • For high-efficiency dialysis, blood flow >300 mL/min is required to achieve target spKt/V of 1.2 4
  • 15% of catheter treatments have blood flow <300 mL/min, leading to underdialysis 4

Balancing Blood Flow with Dialysate Flow

Maintain a blood flow to dialysate flow ratio of 1:2 for optimal efficiency 5:

  • If dialysate flow must be reduced (e.g., from 500 to 400 mL/min), calculate the appropriate blood flow using the K₀A equation to maintain equivalent dialysis efficiency 5
  • Example: When reducing dialysate flow from 500 to 400 mL/min with blood flow of 200 mL/min, increase blood flow to 210 mL/min to maintain clearance 5

Catheter-Specific Considerations

Central venous catheters may require blood flow adjustment based on position and design 4:

  • Femoral catheters have significantly greater recirculation than internal jugular catheters (13.1% vs 0.4%) 4
  • Femoral catheters <20 cm have recirculation of 26.3% vs 8.3% for those >20 cm 4
  • Newer catheter designs can achieve 400 mL/min or greater when properly placed 4
  • Exception: Pediatric or smaller adult catheters are not designed for flows >300 mL/min 4

Common Pitfalls to Avoid

Do not sample immediately at time zero (blood pump still running at full speed): This causes major BUN reduction from recirculated blood still in the access, falsely elevating Kt/V 1

Do not assume that reducing blood flow will improve Kt/V in patients with low access flow: Studies demonstrate that even with compromised access, higher blood flow rates generally increase delivered dose 2

Do not reduce blood flow as the primary solution for inadequate dialysis: The prescribed dose should be Kt/V 1.3 (URR ~70%) to ensure minimum delivered dose of 1.2 is achieved 1. When blood flow is limited, extend treatment time instead 1, 3

Monitor for factors that compromise effective blood flow 1:

  • Access recirculation reducing concentration gradient
  • Blood pump/dialysate flow calibration errors
  • Dialyzer clotting during treatment
  • High pre-pump negative pressure causing tubing collapse

References

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Research

Optimizing dialysis dose by increasing blood flow rate in patients with reduced vascular-access flow rate.

American journal of kidney diseases : the official journal of the National Kidney Foundation, 2001

Guideline

Maximum Blood Flow Rate for Hemodialysis Central Venous Catheters

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Research

Effects of reduced dialysis fluid flow in hemodialysis.

Journal of Nippon Medical School = Nippon Ika Daigaku zasshi, 2013

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.