What causes an Endotracheal Tube (ETT) cuff leak?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: December 26, 2025View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

Causes of Endotracheal Tube (ETT) Cuff Leak

ETT cuff leaks result from either structural defects in the cuff/inflation system or functional problems related to cuff positioning, pressure management, and patient-tube size mismatch. 1

Structural Causes

Cuff and Inflation System Damage

  • Physical trauma to the cuff, pilot balloon, or inflation tubing from inadvertent cutting during ETT repositioning or securing (particularly when removing adhesive tape or ties) can sever the inflation system 2, 1
  • Manufacturing defects in the ETT cuff or inflation system may cause leaks, though this is less common 1
  • The cuff material itself can develop structural defects that allow air to escape 1

Functional Causes (Intact Cuff)

Inadequate Cuff Pressure Management

  • Cuff underinflation is a primary cause of leaks around structurally intact cuffs 1
  • Using the minimal leak test technique for cuff maintenance results in inappropriate pressures in 56% of patients: 24% have pressures <20 cmH₂O (underinflated) and 31% have pressures ≥30 cmH₂O (overinflated) 3
  • Only 44% of patients maintained with the minimal leak test achieve the target pressure range of 20-30 cmH₂O 3

ETT Malposition

  • Cephalad migration of the ETT (partial tracheal extubation) creates a leak as the cuff moves toward the larger diameter upper airway 1
  • Misplaced orogastric or nasogastric tubes can interfere with cuff seal 1

Patient-Tube Size Mismatch

  • Wide discrepancy between ETT diameter and tracheal diameter prevents adequate seal even with proper cuff inflation 1
  • Female gender and obesity are associated with requiring less cuff volume, suggesting anatomical variations affect seal 3

Ventilator Settings

  • Increased peak airway pressure can overcome cuff seal pressure and cause leaks 1
  • Lower PEEP levels are associated with increased cuff leakage (inversely related) 4
  • Peak inspiratory pressure affects leakage, with higher PIP generally reducing leak except at very high PEEP (15 cmH₂O) 4
  • Mode of ventilation influences leak rates, with variation among volume control, pressure control, and CPAP modes 4

Laryngeal Pathology Context

While not a direct cause of cuff leak, it's important to distinguish cuff leak testing (used to detect laryngeal edema) from actual ETT cuff leaks:

  • Laryngeal edema from prolonged intubation (>36 hours) causes reduced air leak around the deflated cuff during cuff leak testing, which predicts post-extubation stridor 5
  • This represents airway pathology rather than ETT malfunction 5

Common Pitfalls

  • Assuming all leaks require ETT replacement: Many leaks resolve with conservative management by correcting cuff pressure, repositioning the ETT, or adjusting ventilator settings 1
  • Relying solely on minimal leak test: This technique is unreliable and should be replaced with cuff manometry to maintain proper pressures of 20-30 cmH₂O 3, 6
  • Ignoring ventilator parameter effects: Adjusting PEEP and PIP can significantly reduce leakage without changing the ETT 4

References

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Guideline

Cuff Leak Test Procedure and Interpretation

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.