Is a Bone-Anchored Hearing Aid (BAHA) Sentio placement medically necessary for a patient with congenital right-sided hearing loss and moderate to severe mixed hearing loss?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: December 29, 2025View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

BAHA Sentio Medical Necessity Assessment

The BAHA Sentio placement is medically necessary for this patient with congenital right-sided congenital malformation and moderate to severe mixed hearing loss, as she meets the anatomical criteria (congenital malformation of the tympanic membrane preventing conventional hearing aid use), but medical necessity cannot be definitively confirmed without verifying that her bone conduction thresholds meet the specific audiologic criteria for the Sentio device.

Critical Issue: Missing Device-Specific Audiologic Criteria

The primary barrier to approval is determining whether the BAHA Sentio's bone conduction threshold requirements align with this patient's audiometric data. The contract policy clearly states that different BAHA models have different bone conduction threshold limits 1:

  • Standard devices (BAHA Attract, Divino, BP100, Baha 4): ≤45 dB HL
  • Power devices (BAHA 5 Power, Intenso, Ponto Plus Power): ≤55 dB HL
  • Super power devices (BAHA Cordelle II): ≤65 dB HL

The BAHA Sentio is not listed in the policy's device specifications, which creates the documentation gap.

Patient Meets Anatomical/Medical Criteria

This patient clearly satisfies the medical necessity requirements for BAHA candidacy 2, 3:

  • Congenital malformation present: Documented malformed tympanic membrane without normal landmarks, status post childhood tympanoplasty for congenital malformation 2
  • Conventional hearing aid contraindicated: The malformed TM and failed surgical reconstruction prevent effective use of air-conduction hearing aids 1, 4
  • Age requirement met: Patient is adult (BAHA approved for ages ≥5 years) 2, 3

Audiologic Data Analysis

Right ear findings (affected side):

  • Pure tone average: 75 dB HL (air conduction)
  • Moderate to severe mixed hearing loss documented
  • Type As tympanogram (shallow compliance, consistent with TM abnormality)
  • Word discrimination: 100% (indicating intact cochlear function)

Left ear findings (better ear):

  • Pure tone average: 30 dB HL
  • Mild high-frequency sensorineural hearing loss
  • Normal tympanogram (Type A)
  • Word discrimination: 100%

Critical missing data: The bone conduction thresholds at 0.5,1,2, and 3 kHz are not explicitly calculated as a pure tone average in the documentation provided. The policy requires bone conduction PTA ≤45,55, or 65 dB HL depending on device model 1, 3.

Device Selection Considerations

The American Academy of Otolaryngology-Head and Neck Surgery recognizes that osseointegrated bone conductive devices effectively transmit sound from the affected side to the better hearing cochlea in cases of single-sided deafness and unilateral conductive/mixed loss 5. Research demonstrates:

  • Unilateral BAHA success: Patients with unilateral conductive or mixed hearing loss show significant improvement in hearing handicap scores and quality of life with BAHA 4
  • Mixed hearing loss outcomes: BAHA achieves air-bone gap closure in 85% of cases with mixed hearing loss 3
  • Congenital atresia/malformation: BAHA represents the treatment of choice for congenital ear malformations, superior to surgical reconstruction 2

Quality of Life Impact

This patient demonstrates significant hearing handicap that warrants intervention 5:

  • Unilateral hearing loss creates substantial communication difficulties even with one normal ear 4
  • 86% of adults with unilateral hearing loss report hearing handicap on standardized inventories 5
  • Associated tinnitus (present in this patient) compounds the disability 5
  • Progressive worsening reported by patient indicates deteriorating function

Recommendation for Authorization Decision

To approve this request, the following must be verified:

  1. Obtain bone conduction pure tone average: Calculate the average of bone conduction thresholds at 0.5,1,2, and 3 kHz for the right ear from the audiogram 1, 3

  2. Match device to thresholds:

    • If BC PTA ≤45 dB HL: Standard BAHA devices appropriate
    • If BC PTA 46-55 dB HL: Power devices required
    • If BC PTA 56-65 dB HL: Super power devices required
    • If BC PTA >65 dB HL: BAHA not indicated; consider cochlear implant evaluation
  3. Determine Sentio classification: Contact the manufacturer or review FDA labeling to establish whether the BAHA Sentio is a standard, power, or super power device and its approved bone conduction threshold range 1

Common Pitfalls to Avoid

  • Do not deny based solely on "not on the list": The Sentio may be a newer model equivalent to listed devices; verify specifications rather than automatically denying 1
  • Do not confuse air conduction with bone conduction thresholds: The 75 dB HL right ear threshold is air conduction; bone conduction determines BAHA candidacy 2, 3
  • Do not overlook the bilateral asymmetry: This patient has asymmetric sensorineural hearing loss between ears, which may affect device selection 6
  • Do not delay unnecessarily: The patient has already failed conventional treatment (tympanoplasty); prolonged delay in amplification affects quality of life 5, 4

Next Steps for Case Manager

Request the following from the provider 1, 3:

  • Bone conduction pure tone average (0.5,1,2,3 kHz) for right ear
  • BAHA Sentio manufacturer specifications confirming bone conduction threshold range
  • Confirmation that conventional air-conduction hearing aids have been considered and deemed inappropriate due to the anatomical malformation

If bone conduction thresholds are within device specifications and the Sentio is FDA-approved for those thresholds, approve the request as medically necessary 2, 3, 4.

References

Research

Evolving applications in the use of bone-anchored hearing aids.

American journal of audiology, 2002

Research

Bone-anchored hearing aids (Baha) in congenital aural atresia: personal experience.

International journal of pediatric otorhinolaryngology, 2011

Research

Results and complications of the Baha system (bone-anchored hearing aid).

European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery, 2010

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Research

The Baha system in patients with single-sided deafness and contralateral hearing loss.

Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery, 2010

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.