What are the appropriate settings for an electrocardiogram (ECG)?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: January 15, 2026View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

Standard ECG Technical Settings

The standard ECG should be recorded at 10 mm/mV calibration (where 1 mV produces a 10 mm deflection), with a paper speed of 25 mm/second, using a high-frequency cutoff of at least 150 Hz for adults and adolescents, and 250 Hz for infants and children. 1

Core Calibration Parameters

Amplitude (Gain) Settings

  • Standard calibration is 10 mm/mV universally, meaning each small box (1 mm) represents 0.1 mV and each large box (5 mm) represents 0.5 mV 1
  • Alternative gain settings (5 mm/mV for high-amplitude complexes or 20 mm/mV for low-amplitude complexes) may be used but must be clearly documented on the tracing, as this fundamentally alters voltage-based diagnostic criteria 1

Frequency Response Requirements

For Adults and Adolescents (>16 years):

  • High-frequency cutoff: minimum 150 Hz to accurately capture rapid QRS upstrokes and small deflections 2, 1
  • Low-frequency cutoff: 0.05 Hz for routine filters, or relaxed to 0.67 Hz for linear digital filters with zero phase distortion 1

For Infants and Children:

  • High-frequency cutoff: 250 Hz required to capture the more rapid deflections characteristic of pediatric ECGs 2, 1
  • This higher bandwidth prevents amplitude underestimation and missed small deflections 2

Digital Sampling Specifications

  • Minimum sampling rate: 500 samples per second to allow accurate 150 Hz filtering and reduce amplitude measurement errors to <1% 1
  • A/D conversion: minimum 12-bit precision for adequate signal resolution 1
  • Noise levels should be reduced to <1 µV (root mean square) in the averaged signal 1

Equipment and Recording Standards

Electrocardiographic System Requirements

  • The system must meet American Heart Association specifications for frequency response and signal fidelity 2
  • Silver-silver chloride electrodes are recommended as most dependable for minimizing motion artifact 2
  • Continuous oscilloscopic monitoring of minimum 3 leads is recommended, though 12-lead printed capability is highly recommended for accurate interpretation 2

Critical Technical Considerations

  • Use the largest-amplitude deflection in each lead as representative of that measurement, per ANSI/AAMI standards 1
  • Digital electrocardiographs should form "templates" from averaged beats to reduce beat-to-beat variability, which significantly improves measurement reproducibility compared to single-beat analysis 2
  • Electrocardiographs should automatically alert users when suboptimal high-frequency cutoff (such as 40 Hz) is used, as this invalidates amplitude measurements 2

Environmental Conditions

Room Temperature and Humidity

  • Maintain temperature range of 20°C to 22°C (68°F to 71.6°F) for optimal patient comfort and stable cardiovascular responses 2
  • Keep humidity below 60% to prevent variable cardiovascular responses and maintain maximum performance 2
  • Use circulating fans to assist temperature control and ventilation 2

Barometric Considerations

  • If performing gas exchange measurements concurrently, measure barometric pressure and temperature, as gases expand with heat/low pressure and contract with cold/high pressure 2
  • Most automated cardiopulmonary testing systems adjust for ambient conditions automatically 2

Common Pitfalls to Avoid

Setting high-frequency cutoff to 40 Hz to reduce noise will invalidate all amplitude measurements used for diagnostic classification, rendering voltage criteria for left ventricular hypertrophy and other diagnoses unreliable 2

Inadequate skin preparation is a frequent source of poor signal quality regardless of electrode type—extra time spent on preparation results in more stable recordings 2

Using Mason-Likar limb lead placement without obtaining a standard 12-lead ECG first may alter inferior lead complexes to either mimic or hide previous Q waves 2

Failing to document non-standard calibration when using 5 mm/mV or 20 mm/mV settings makes the tracing uninterpretable for voltage-based criteria 1

Interpretation Requirements

All computer ECG interpretations must be verified by a qualified physician, particularly interpretations of rhythm disturbances, ischemia, or infarction 2

Computer programs can provide accurate heart rate, intervals, and electrical axes, but a skilled physician must integrate clinical data, compare with previous tracings, and recognize that identical ECG patterns may occur in different pathophysiologic states 2

References

Guideline

ECG Calibration and Technical Requirements

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.