What does a +3.6 change in bone mineral density from a prior Dual-Energy X-ray Absorptiometry (DEXA) scan indicate in an older adult patient?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: January 21, 2026View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

Understanding a +3.6 Change in DEXA Bone Mineral Density

A +3.6 change in bone mineral density from a prior DEXA scan is highly unusual and almost certainly represents a measurement error, technical artifact, or reporting inconsistency rather than a true physiological change in bone density.

Why This Change is Implausible

The magnitude of this change defies biological plausibility. To understand why, we need to clarify what type of measurement this represents:

If This Represents a T-score or Z-score Change:

  • A change of +3.6 standard deviations in bone mineral density between scans is physiologically impossible 1
  • Normal bone density is defined as a T-score greater than -1.0, osteopenia as -1.0 to -2.4, and osteoporosis as ≤-2.5 1
  • Even the most aggressive bone-building treatments (bisphosphonates, denosumab, teriparatide) typically increase BMD by only 3-8% over 1-2 years, which translates to approximately 0.3-0.8 standard deviation improvement at most 2
  • Studies in cancer survivors show that BMD Z-scores increase modestly over time (2-10 years post-treatment), but never by multiple standard deviations 2

If This Represents a Percentage Change:

  • A +3.6% change in BMD is within the realm of biological possibility but still requires careful interpretation
  • The least significant change (LSC) for DEXA machines—the minimum change needed to exceed measurement error—typically ranges from 2-4% depending on the skeletal site and machine precision 1, 3
  • A +3.6% increase could represent a true treatment response if the patient has been on bisphosphonates or other bone-building therapy for 1-2 years 2

Common Technical Errors to Investigate

Before accepting this result as valid, systematically evaluate these potential sources of error:

Scanner and Technical Factors:

  • Different DEXA machines or scan modes produce significantly different BMD measurements 4
  • Hologic machines show statistically significant differences between Array, Fast Array, and Express Array modes (differences of 0.035-0.037 g/cm²) 4
  • Patients must be scanned on the same machine for accurate longitudinal comparison 1, 4
  • Cross-calibration errors between different manufacturers or models can produce spurious changes 4

Positioning and Analysis Errors:

  • Improper patient positioning is one of the most common sources of measurement error 3
  • Incorrect region of interest (ROI) placement during scan analysis can dramatically alter results 3
  • Vertebral degenerative changes, osteophytes, and aortic calcification artificially elevate lumbar spine BMD measurements, particularly in elderly patients 2, 3
  • Errors in demographic information (age, sex, ethnicity) will change the reference database and alter T-scores and Z-scores 3

Site-Specific Considerations:

  • Lumbar spine measurements are unreliable in elderly patients due to degenerative changes 2
  • Studies show DEXA can overestimate BMD compared to quantitative CT due to abdominal aortic calcifications 2
  • Different skeletal sites (lumbar spine, femoral neck, total hip) show different rates of change and measurement precision 5, 3

Recommended Action Steps

Immediately take the following steps to clarify this result:

  1. Verify the scan was performed on the same DEXA machine as the prior scan 1, 4

    • If different machines were used, the comparison is invalid
    • Request cross-calibration data if available
  2. Review both scan reports in detail 3:

    • Confirm the same skeletal sites were measured (L1-L4 spine, femoral neck, total hip)
    • Verify the same scan mode was used (Array vs Fast Array vs Express Array for Hologic machines) 4
    • Check that demographic information is identical on both scans
    • Examine the actual BMD values in g/cm² rather than just T-scores or Z-scores
  3. Assess for analysis errors 3:

    • Review the region of interest placement on both scans
    • Look for vertebral compression fractures, hardware, or degenerative changes that might affect measurements
    • Consider whether lateral projection might be more appropriate than PA projection 5
  4. If the change appears valid, evaluate for secondary causes 2:

    • Review medication history (bisphosphonates, denosumab, teriparatide, hormone replacement)
    • Assess for resolution of conditions causing bone loss (vitamin D deficiency, hypogonadism, hyperthyroidism)
    • Consider whether the patient has gained significant weight (BMI affects BMD measurements)
  5. Repeat the DEXA scan 2, 1:

    • On the same machine as the most recent scan
    • With careful attention to positioning and analysis
    • If results remain discordant, consider alternative imaging (quantitative CT) for clarification 2

Clinical Context Matters

The clinical plausibility depends entirely on the patient's treatment history:

  • Untreated patients or those on calcium/vitamin D alone: A +3.6% change might represent measurement error or the upper limit of natural variation 2
  • Patients on bisphosphonates for 1-2 years: A +3.6% increase is plausible and represents a good treatment response 2
  • Patients with previously untreated vitamin D deficiency now repleted: Modest improvements are expected, but not dramatic changes 2

The most likely explanation remains technical error until proven otherwise through systematic verification of the measurement methodology 4, 3.

References

Guideline

Osteoporosis Screening with DEXA Scans

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Research

Fundamentals and pitfalls of bone densitometry using dual-energy X-ray absorptiometry (DXA).

Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA, 2004

Research

Comparison of Differences in Bone Mineral Density Measurement With 3 Hologic Dual-Energy X-Ray Absorptiometry Scan Modes.

Journal of clinical densitometry : the official journal of the International Society for Clinical Densitometry, 2021

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.