How can anesthetic gas machine vaporizers be classified according to characteristics such as Concentrated-calibrated, Measured flow, Flow-power, Bubble through, and Injection?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: January 22, 2026View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

Classification of Anesthetic Gas Machine Vaporizers

Anesthetic vaporizers can be classified into five main categories based on their operational characteristics: concentration-calibrated (variable bypass), measured flow, flow-power, bubble-through, and injection systems, with modern practice predominantly utilizing concentration-calibrated vaporizers due to their superior safety profile and accuracy.

Concentration-Calibrated (Variable Bypass) Vaporizers

  • These are the most common vaporizers in modern anesthetic practice, featuring a dial that directly displays the output concentration of volatile agent 1
  • The design splits incoming fresh gas flow into two streams: one passes through the vaporizing chamber (picking up anesthetic vapor), while the bypass stream flows around it, with the two streams mixing at the outlet to achieve the dialed concentration 1
  • Modern concentration-calibrated vaporizers are flow and temperature compensated, meaning they maintain accurate output concentrations despite variations in fresh gas flow rates and ambient temperature 1
  • These vaporizers are agent-specific and must be filled only with their designated anesthetic agent to prevent dangerous concentration errors 2
  • Safety features include anti-spill mechanisms, select-a-tec interlocking systems (preventing simultaneous activation of multiple vaporizers), and agent-specific filling devices 3, 1

Measured Flow Vaporizers

  • In measured flow systems, a known flow of carrier gas passes through the vaporizing chamber, and the output concentration is calculated based on the measured flow rate through the chamber 4
  • The Aladin cassette vaporizer represents a modern measured flow system that can function as both a variable bypass and measured flow vaporizer, integrated with specific anesthesia machines 4
  • These systems require electronic control units within the anesthesia machine and use agent-specific detachable cassettes as vaporizing chambers 4
  • A critical limitation is that measured flow vaporizers require electrical power to function, making them dependent on machine electronics 4

Flow-Power (Electronically Controlled) Vaporizers

  • Flow-power vaporizers utilize both electronic and pneumatic control to deliver precise agent concentrations, particularly effective with low fresh gas flows 4
  • Desflurane requires a specialized heated, pressurized vaporizer due to its unique physical properties (boiling point near room temperature at 23.5°C), which necessitates differential pressure sensing between saturated desflurane flow and fresh gas flow 5
  • Modern flow-power systems are controlled by central processing units that continuously monitor and adjust vapor concentration by altering fresh gas flow through the vaporizer 1
  • These systems can achieve rapid changes in end-tidal agent concentration, even at metabolic flow rates, allowing quick alterations in anesthetic depth with minimal waste 4

Bubble-Through Vaporizers

  • Bubble-through designs force carrier gas through a liquid anesthetic agent, creating bubbles that become saturated with vapor 6
  • These are older, less precise systems that lack the temperature and flow compensation of modern vaporizers, making their output concentration highly variable 6
  • The output concentration depends heavily on factors such as fresh gas flow rate, temperature of the liquid agent, and the surface area of gas-liquid interface 6
  • Due to their unpredictability and safety concerns, bubble-through vaporizers are rarely used in modern anesthetic practice 6

Injection Vaporizers

  • Injection systems directly inject liquid anesthetic into the breathing circuit, where it vaporizes due to the heat and humidity of the patient's exhaled gases 4
  • Two main types exist: the Maquet injection vaporizer (customized for Maquet FLOW-i machines) and the DIVA system (designed for Dräger Zeus machines) 4
  • These systems can achieve set end-tidal agent concentrations very rapidly, even with metabolic flow rates, resulting in minimal anesthetic wastage and reduced operating room pollution 4
  • The AnaConDa device combines a vaporizer with a heat and moisture exchange filter, designed primarily for intensive care sedation and out-of-operating-room use 4
  • Injection systems require simultaneous sensing of inhaled and exhaled gas concentrations and are typically used in closed-circuit breathing systems 5

Critical Safety Considerations Across All Vaporizer Types

  • Always verify that vaporizers are correctly fitted with locking mechanisms fully engaged and that control knobs rotate through their complete range 7, 8
  • Check that vaporizers are adequately filled but not overfilled, with filling ports tightly closed 7, 8
  • Never tilt vaporizers during filling or transport, as this can result in dangerously high vapor concentrations 7, 8
  • For non-interlocked systems, never operate multiple vaporizers simultaneously, as this can produce uncontrolled and potentially lethal anesthetic concentrations 3
  • Manual leak testing should only be performed on basic Boyle's machines, as it may harm modern anesthetic workstations—always consult manufacturer recommendations 7, 8
  • When changing vaporizers during use, repeat leak testing whenever possible, as failure to do so is a common cause of critical incidents 8

References

Research

Modern anaesthesia vapourisers.

Indian journal of anaesthesia, 2013

Guideline

Safety Considerations for Non-Interlocked Vaporizers

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Research

Advances in vaporisation: A narrative review.

Indian journal of anaesthesia, 2020

Research

Advanced technologies and devices for inhalational anesthetic drug dosing.

Handbook of experimental pharmacology, 2008

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Guideline

Vaporiser Safety and Use Guidelines

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Related Questions

How should vaporisers in anaesthesia machines be selected and used?
What are the safety considerations when using non-interlocked (vaporizers that do not prevent simultaneous administration of multiple anesthetic agents) vaporizers?
What is the recommended sequence of arrangement for non-interlocked vaporizers (anesthetic delivery devices)?
What is the ideal splitting ratio in a vaporizer for safe and effective delivery of anesthetic vapor?
What is the recommended dose of Alteplase (tissue plasminogen activator) for an adult patient with acute ischemic stroke?
What is the recommended management for an elderly patient with a history of osteoporosis (bone disease characterized by low Bone Mineral Density (BMD)) on chronic prednisone (glucocorticoid) therapy?
What is the recommended treatment for a patient with tonsillitis, considering their age, weight, and medical history, including any allergies or previous reactions to medications?
What are the risk factors for developing fungal infections, particularly in individuals with weakened immune systems, such as those with Human Immunodeficiency Virus (HIV)/Acquired Immunodeficiency Syndrome (AIDS), diabetes, or those taking immunosuppressive medications like prednisone?
What is the etiology, clinical presentation, diagnosis, findings, commonly affected organs, and treatment for pediatric gunshot wounds, including damage control?
Can gas pain cause right lower quadrant pain in patients with a history of gastrointestinal issues such as irritable bowel syndrome (IBS), inflammatory bowel disease (IBD), or gastroparesis?

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.