How to Calculate Light's Criteria for Pleural Effusion
Light's criteria classify a pleural effusion as an exudate if at least one of three specific ratios is met: pleural fluid to serum protein ratio >0.5, pleural fluid to serum LDH ratio >0.6, or pleural fluid LDH >0.67 (two-thirds) of the upper limit of normal serum value. 1
Required Laboratory Values
To calculate Light's criteria, you need the following measurements:
- Pleural fluid protein and serum protein (measured simultaneously) 1
- Pleural fluid LDH and serum LDH (measured simultaneously) 1
- Upper limit of normal for serum LDH from your laboratory's reference range 1
Step-by-Step Calculation
Calculate the following three ratios:
Protein ratio: Divide pleural fluid protein by serum protein 1
- If >0.5 → suggests exudate
LDH ratio: Divide pleural fluid LDH by serum LDH 1
- If >0.6 → suggests exudate
Absolute pleural fluid LDH: Compare to upper limit of normal serum LDH 1
- If pleural fluid LDH >0.67 (or two-thirds) of the upper normal serum limit → suggests exudate
Interpretation
- If ANY ONE of these three criteria is met, classify the effusion as an exudate 1
- If NONE of the criteria are met, classify as a transudate 1
Performance Characteristics
- Light's criteria have excellent sensitivity (98%) for detecting exudates, meaning they rarely miss an exudative effusion 1
- However, specificity is lower (72%), meaning approximately 25-30% of transudates (especially cardiac and hepatic) may be misclassified as exudates 2
- The positive likelihood ratio is 3.5 and negative likelihood ratio is 0.03 1
Alternative When Serum Samples Unavailable
If you cannot obtain serum samples, use these pleural fluid-only criteria:
- Pleural fluid LDH >200 IU/L (or >67% of upper normal serum limit) 2, 3
- Pleural fluid cholesterol >55 mg/dL 2
Important Caveats
- Analytical platform variability: Different laboratory analyzers can produce discordant results in up to 18% of cases, particularly affecting LDH measurements 4
- Diuretic effect: Patients with heart failure on diuretics may have transudates misclassified as exudates by Light's criteria 5
- Borderline protein values: Light's criteria are most useful when pleural fluid protein is 25-35 g/L, where classification is uncertain 4
- Consider serum-effusion albumin gradient (SEAG): If Light's criteria suggest exudate but clinical picture suggests transudate (e.g., heart failure), SEAG has 97.5% accuracy for identifying transudates 5