How is dual calibration of an electrocardiogram performed and what are the standard voltage (10 mm/mV) and paper speed (25 mm/s) settings?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: February 6, 2026View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

ECG Dual Calibration: Standard Settings and Performance

Standard ECG calibration is universally set at 10 mm/mV amplitude and 25 mm/s paper speed, where 1 mV produces a 10 mm vertical deflection and each 1 mm horizontal box represents 0.04 seconds. 1

Standard Calibration Parameters

Amplitude (Voltage) Calibration

  • Standard gain: 10 mm/mV where each small box (1 mm) represents 0.1 mV and each large box (5 mm) represents 0.5 mV 1
  • This setting is universally recommended for routine ECG recordings across all age groups 1
  • Alternative calibrations (5 mm/mV for high-amplitude signals or 20 mm/mV for low-amplitude signals) must be clearly documented on the tracing, as they invalidate standard voltage-based diagnostic criteria 1

Paper Speed Calibration

  • Standard paper speed: 25 mm/s where each small box (1 mm) represents 0.04 seconds and each large box (5 mm) represents 0.20 seconds 1
  • At this speed, accurate temporal measurements of intervals (PR, QRS, QT) can be performed 1

How Dual Calibration is Performed

Calibration Signal Display

  • A standardization mark (calibration pulse) should be visible on the ECG tracing, typically appearing as a square wave deflection at the beginning or end of the recording 2
  • This calibration pulse verifies that the machine is properly calibrated before interpretation begins 2
  • When obtaining waveform strips for documentation, verify that voltage standardization and paper speed are visible on the strip 2

Technical Requirements for Accurate Calibration

Digital Sampling and Filtering Standards:

  • Minimum sampling rate of 500 samples per second is required to allow accurate 150 Hz filtering and reduce amplitude measurement errors to <1% 2, 1
  • A/D conversion requires minimum 12-bit precision for adequate signal resolution 1

Frequency Response Requirements:

  • For adults and adolescents: high-frequency cutoff of at least 150 Hz 2, 1
  • For infants and children: high-frequency cutoff of 250 Hz to capture rapid deflections accurately 2, 1
  • Low-frequency cutoff of 0.05 Hz for routine filters, or relaxed to 0.67 Hz for linear digital filters with zero phase distortion 2, 1

Critical Pitfalls to Avoid

Technical Errors That Invalidate Calibration

  • Suboptimal high-frequency cutoff (such as 40 Hz used for noise reduction) invalidates all amplitude measurements used for diagnostic classification 2
  • Electrocardiographs should automatically alert users when suboptimal settings are used and restore proper settings between recordings 2
  • Inadequate high-frequency response systematically underestimates QRS amplitude and smooths important features like Q waves 2

Documentation Requirements

  • Always verify the calibration settings are displayed on printed or saved ECG strips 2
  • Non-standard calibration settings must be explicitly documented to prevent misinterpretation of voltage-dependent criteria (such as LVH criteria) 1
  • When using electronic health records, ensure voltage standardization is visible when waveforms are scanned or electronically transferred 2

Age-Specific Calibration Considerations

  • Adults >35 years: Standard 10 mm/mV calibration with 150 Hz high-frequency cutoff 1
  • Adolescents (16-35 years): Standard calibration, but recognize voltage criteria for LVH have lower accuracy in this age group 1
  • Children and infants: Standard 10 mm/mV calibration but requires 250 Hz high-frequency cutoff to capture rapid deflections accurately 2, 1

Quality Assurance Measures

  • Use the largest-amplitude deflection in each lead as representative of that measurement, per ANSI/AAMI standards 1
  • Noise levels should be reduced to <1 µV (root mean square) in the averaged signal 1
  • Digital electrocardiographs improve reproducibility by creating representative templates from accurately aligned complexes 2
  • Verify electrode placement accuracy before attributing low voltage to pathology, as misplacement is the most common cause of apparent calibration errors 3

References

Guideline

ECG Calibration and Technical Requirements

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Guideline

EKG Interpretation Guidelines

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.