What are the ECG characteristics of ventricular pacing?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: February 14, 2026View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

Ventricular Pacing ECG Characteristics

Ventricular pacing produces a characteristic ECG pattern with a pacing spike followed by a wide QRS complex (>120 ms) that typically shows a left bundle branch block (LBBB) morphology with left axis deviation when pacing from the traditional right ventricular apex. 1

Essential ECG Features of Ventricular Pacing

Basic Morphology

  • Pacing artifact (spike) appears as a sharp vertical deflection immediately before the QRS complex, representing the electrical impulse delivered by the pacemaker 2, 3
  • Wide QRS complex (>120 ms duration) follows each pacing spike when capture is successful, reflecting abnormal ventricular depolarization from the pacing site rather than through the normal conduction system 1, 4
  • QRS-T wave discordance is the anticipated pattern, where the ST segment and T wave deflect opposite to the main QRS vector 4

Right Ventricular Apical Pacing (Traditional Site)

  • LBBB configuration with predominantly negative QRS in V1 and positive in V6 1
  • Left axis deviation in the frontal plane 1
  • This pattern occurs because electrical activation spreads from the RV apex across the septum to activate the left ventricle in a delayed, abnormal sequence 1

Left Ventricular Pacing

  • Right bundle branch block (RBBB) configuration with positive QRS in V1 3, 1
  • Axis depends on lead position within the left ventricle 1
  • Absence of positive R-wave in V1 during intended biventricular pacing may indicate LV lead displacement or loss of LV capture 3

Verifying Proper Capture

ECG Confirmation

  • Each pacing spike must be followed by a corresponding QRS complex (for ventricular pacing) with appropriate morphology 3
  • Absence of QRS after pacing spike indicates failure to capture, meaning the electrical impulse failed to depolarize the myocardium 2
  • Optimize ECG lead selection to minimize pacemaker artifact and maximize QRS visualization 3

Mechanical Confirmation (Critical)

  • Never rely on ECG alone - always verify both electrical and mechanical capture 2
  • Confirm peripheral pulse through arterial line waveform, pulse oximeter plethysmogram, or manual palpation to ensure each pacing stimulus produces mechanical systole 3
  • Verify adequate blood pressure and perfusion to confirm effective mechanical capture 3

Common Pacemaker Modes and ECG Patterns

VVI/VVIR Mode

  • Paces and senses only in the ventricle 5
  • Pacing spike followed by wide QRS when pacing occurs 5
  • No atrial activity is tracked or paced 5

DDD/DDDR Mode

  • May show atrial pacing spikes followed by P waves and/or ventricular pacing spikes followed by QRS complexes 6
  • Maintains AV synchrony when functioning properly 6

Algorithms Minimizing Ventricular Pacing

  • May produce "bizarre" ECG appearances that can suggest malfunction but are actually normal algorithm function 6
  • Can show longer AV delays for sensed events than paced events, violating traditional AV conduction rules 6
  • Mode conversion algorithms switch between AAI(R) and DDD(R) based on AV conduction, creating variable ECG patterns 6

Critical Pitfalls to Avoid

Failure to Capture

  • Pacing spike without QRS represents failure to capture and requires immediate intervention, especially in pacemaker-dependent patients 2
  • Temporary transvenous pacemakers have higher risk of capture loss due to lead dislodgement, occurring in up to 5.7% of cases with 50% within first 24 hours 3

Distinguishing from Other Malfunctions

  • Undersensing: Pacing spikes occur inappropriately when intrinsic QRS complexes are present, but spikes still produce QRS complexes 2
  • Oversensing: Inappropriate inhibition of pacing with fewer spikes than expected because the pacemaker misinterprets noise as cardiac activity 2
  • Failure to pace: Complete absence of pacing spikes when needed 2

Misinterpreting ST-T Changes

  • The altered intraventricular conduction from ventricular pacing changes ST-segment and T-wave configuration 4
  • Expected pattern is QRS-ST-T discordance, not concordance 4
  • Diagnosing acute myocardial infarction requires recognizing deviations from this expected discordant pattern 4

Monitoring Recommendations

  • Continuous ECG monitoring for 12-24 hours post-implantation to detect early lead dislodgement 3
  • All patients with temporary transvenous pacing require continuous monitoring until device removal or permanent system placement 3

References

Guideline

Pacemaker Spike Without QRS: Failure to Capture

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Guideline

Determining Good Capture in a Paced Rhythm

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2025

Research

The electrocardiographic diagnosis of acute myocardial infarction in patients with ventricular paced rhythms.

Academic emergency medicine : official journal of the Society for Academic Emergency Medicine, 1998

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.

Have a follow-up question?

Our Medical A.I. is used by practicing medical doctors at top research institutions around the world. Ask any follow up question and get world-class guideline-backed answers instantly.